4.1 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN Stretcher Siderail
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 7
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.1
PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/
OB/GYN Stretcher Siderail
Tools required:
Ratchet
½" socket
½" open end wrench
Torque wrench 0-250 in-lb
Removal
1. Lower the siderail.
2. Using the ½" open end wrench, the ratchet, and the ½" socket, remove the
two swing pivot nuts (C) and washers (E) from the swing pivot bolts (A)
that secure the siderail to the swing arms (D) (see figure 4-1 on page 4-7)
or (see figure 4-2 on page 4-8).
Figure 4-1. PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical Stretcher
Siderail
D
m115a010