3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 15
Calibration Parameters
With the calibration password entered the different parameters of the instrument can
be calibrated:
To enter the different parameters, complete the following procedure:
1. Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
e.g. ‘DC Volts’ and then press
SELECT
soft key
Depending upon the model and options fitted the available functions will vary