25
13. After entering the exact value, press the NET/GROSS key to save the value. If the cali-
bration was successful, the display will show "
EndC1
" momentarily, followed by "
C 2
" for
the second calibration point.
14. If using multi-point calibration, repeat steps 10-11 for C2 and C3.
Otherwise, enter a
zero value in for C2
. At the conclusion of this step, the indicator displays the current
gravity settings, e.g. “9.800”.
15. If the local gravity factor is known, then use the up/down and left/right keys to adjust the
displayed value. Otherwise, just press the NET/GROSS key. The display will
show “-
donE”.
16. Press the NET/GROSS key to revert
up to “F17”.
17. If the calibration was
not
successful, one of the error messages below will appear. Take
the indicated action to correct the problem, and then perform a new calibration.
•
"
Err0
" - The calibration test weight or the keyed-in weight is larger than the full capacity
of the scale. Change the calibration test weight or check the input data.
•
"
Err1
" - The calibration test weight or the keyed-in weight is smaller than 1% of the full
capacity of the scale. Change the calibration test weight or check the input data.
•
"
Err2
"
– There is not enough signal from the load cells to establish a proper calibration.
Most common causes include incorrect load cell wiring, a mechanical obstruction or a
faulty (damaged) load cell.
Key-In Zero Calibration Value (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid zero calibration value, obtained from a successful F16 calibration procedure, must
be used.
1.
While in the Setup mode, scroll to "
F 19
", and then scroll down once using the ZERO key.
The display will momentarily show "
C 0
", followed by a value.
2.
Use the four directional keys to enter in the actual zero calibration value.
3.
After entering the exact value, press the NET/GROSS key to save the value. The display
will revert up to F19.
Key-In Span Calibration Value (F20)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss,
but it can also be used to manually linearize your scale. Valid span calibration values,
obtained from a successful F17 calibration procedure, must be used.
1.
While in the Setup mode, scroll to "
F 20
", and then scroll down once using the ZERO key.
The indicator will prompt you to enter the information in the Table following.
2.
If the value shown is correct, press the NET/GROSS key to move to the next parameter.
Otherwise, use the four directional keys to enter in the actual calibration value
3.
After setting the exact value, press the NET/GROSS key to save the value.
4.
At the completion of the sequence, the indicator will then revert up to F20.