background image

Trafag Italia Srl

Via Cremona 1
20025 Legnano (Mi)

Tel. +39 0331 59 23 97
Fax +39 0331 59 98 15

www.trafagitalia.com
[email protected]

INSTRUCTIONS MANUAL

BIMETAL THERMOMETERS

Instruction H21017a

05/2017

BIMETAL THERMOMETERS

TYPE: TMT501, TMT502

1 GENERAL WARNINGS

1.1 Instruments measuring range

Verify that measuring range of the instrument is suitable with wor-
king conditions of the circuit/plant. Maximum process temperatu-
re must remain within the instrument measuring range, identified 
on dial with two black triangles.

1.2  Admitted over temperatures

Accidental over temperatures are admitted for short periods, provi-
ded that they remain within the values indicated in the datasheet.

1.3  Ambient temperature

Instruments are designed to work with an ambient temperature 
between -20°C/+60°C (-4°F/+140°F). Different ambient tempera-
tures could give wrong indications or to damage the instrument. IT 
is recommended to verify that the instrument installation position 
is not subject to direct heat sources, both for convention or radia-
tion: if this is not possible, please foresee a protective screen. 

1.4  Mechanical vibrations

Verify that chosen position for the installation is not subjected to 
continuous mechanical vibrations, because this will cause accuracy 
errors and shorter working life of instrument. If this is not possi-
ble, it is recommended to use instruments filled with dampening 
liquid.

2 INSTALLATION

2.1  Process installation

The thermometers must be installed in a thermowell that is sui-
table to the line or to the tank where the temperature is to be me-
asured. It is also necessary to verify that the thermowell is suitable 
for the installation of the thermometer. With reference to the figu-
re N° 3 is necessary to verify that:

 

The minimum length “S” (see under table 3) guarantee that the 
sensing element is totally immerged into dimension of ther-
mowell and consequently into the piping, for best sensibility.

 

Bulb diameter 

d

 must be maximum 1mm less than the bore 

diameter of thermowell.

 

Thermowell thread must be suitable with thermometer thread.

Then insert the instrument bulb till the bottom of thermowell, 
screw the rotating connection of thermometer on thermowell 
thread and orient the instrument in a position that permit an easy 
reading. Then tighten with suitable key the rotating nut and verify 
the solidity of the mounting.

NOTE: 

In the event that the tolerance between bulb diameter and 

bore diameter of thermowell is more than 1mm fill the thermowell 
bore with suitable transmission fluid such as silicone oil or graphi-
te dust : this will improve the instrument performances.

3 Regulation of the instrument in 

“every angle” version

Instrument case could be oriented before installation to the requi-
red inclination, up to a maximum of 90°. To do this it is necessary to 
unscrew the two “A” screws, place the case in the required inclina-
tion and then screw again the two “A” screws. (see Fig. 2)

Fig. 2 – Regulation of case inclination

WARNING

 Absolutely do not use the instrument case to tighten it 

on thermowell. This can cause damages to the instrument.

Reviews: