Tradinco TRAQC-3 FM User Manual Download Page 9

User Manual

 

TRAQC-3 FM 

 

 

TRAQC-3 LH  10/2014 EN Rev.2

 

 

 

page : 9 

 
 

2.6.

 

Specifications    

Temperature range  

: A5°C to 375°C. 

Resolution 

 

: 0,1 °C/°F. 

Accuracy 

 

: Over full range better than +/- 0,4 °C, typ. +/- 0.2°C 

Stability 

 

: +/- 0,05 °C. 

Heating time  

 

: Ambient to 300 °C 5.75 minutes. 

Cool down time    

: 300 °C  to 60 °C  14 minutes. 

Stabilization   

 

: 5 minutes. 

Well depth     

 

: Max 100 mm. 

Switch test    

 

: Powerless switch, LED indicator, 5 VDC, 16 mA max. 

Communication    

: RS232 for use with e.g. AutoCal 

Supply voltage      

: 230Volt, 50/60 Hz or 110 Volt, 50/60 Hz.   Selectable. 

Power consumption  

: 250 Watts. 

Fault protection   

  : Internal sensor break detection. 

Fuse 

 

: 3.15 A . 

Thermal Protection switch : 75 °C 
Temp switch test 

  : 5 VDC, 16 mA max. 

Safety  

 

: EN 61010-1:2010/C1:2011, EN61010-2-010:2014. 

Conformity EMC 

: EN-61326-1:2013

 

Weight 

 

: approx 1740 grams, without powercord, incl. 6.2 mm insert. 

Dimensions  

 

: w x d x h : 149 x 159 x 74 mm. 

 
All specifications at 20°C +/- 2°C. 

2.7.

 

Maintenance     

Allow the TRAQC-3 to cool down as described in chapter 2.4. 
Switch the TRAQC-3 off and disconnect the mains plug. 

 
Cleaning TRAQC-3 with inserts: 

A small amount of dust or pollution can cause  jams when the instrument is operated 
with inserts. To prevent this, remove the insert from the heating block at regular 
intervals and if the TRAQC3 is not going to be operated for a longer period. Flush the 
heating block bore with compressed air and clean the bore and inserts with a dry cloth. 

 
Cleaning the fan grille: 

Each TRAQC-3 is fitted with a small air grille via which cooling air enters the TRAQC-3. 
Clean the grille at regular intervals (vacuuming or brushing) depending on the level of air 
pollution. 
 

External cleaning: 

Clean the outside of the TRAQC-3 with a damp cloth and some water or with a 
solvent-free mild cleaning agent. Disconnect the instrument from mains before cleaning. 
 

Fuse: 

Two Fuses of 3.15Amp (slow) need to be used. 

 

Summary of Contents for TRAQC-3 FM

Page 1: ...TRAQC 3 FM Temperature Calibrator User Manual...

Page 2: ...3 Instrument Label 4 1 4 Scope of this manual 4 2 INTRODUCTION 5 2 1 Heating performance 5 2 2 Control Performance 5 2 3 Functions and connections 6 2 4 Operating instructions 7 2 5 Temperature switc...

Page 3: ...charge for parts or labor or will provide a replacement in exchange for the defective product In order to obtain service under this warranty Customer must notify TRADINCO INSTRUMENTS of the defect be...

Page 4: ...ock Attention signal or remark 1 3 Instrument Label 1 4 Scope of this manual This manual contains information for instruments built since September 2008 This manual is a users manual for operating the...

Page 5: ...ith care and only in areas without hazardous fumes or fire risks and get acquainted with all the facilities offered by this field temperature calibrator 2 1 Heating performance The heating block is ma...

Page 6: ...nctions and connections 6 7 8 9 10 11 3 4 5 1 2 1 Heat indicator 2 Up and down push buttons 3 Hole for inserts 4 Switching indicator Led 5 Temp switch terminals 4mm 6 Voltage selector 7 ON OF Switch 8...

Page 7: ...ly XX a code number for manufacturers use only Hr XX The amount of hours the unist has been powered the six digits after the Hr represent the hours up to 99999 9 hrs XXXX see above 4 Place unit under...

Page 8: ...he mass of the probe may influence the measurement considerably Note If the manufacturer has enabled the most left button one has access to a part of a menu structure Use of this feature is not requir...

Page 9: ...ensions w x d x h 149 x 159 x 74 mm All specifications at 20 C 2 C 2 7 Maintenance Allow the TRAQC 3 to cool down as described in chapter 2 4 Switch the TRAQC 3 off and disconnect the mains plug Clean...

Page 10: ...ed and tested with measuring equipment in accordance with recognized national standards prior to delivery The TRAQC 3 should depending on the application situation be inspected at appropriate interval...

Page 11: ...User Manual TRAQC 3 FM TRAQC 3 LH 10 2014 EN Rev 2 page 11 Tradinco Instrumenten Apparaten b v Radonstraat 250 2718 TB Zoetermeer The Netherlands 31 79 2033133 www tradinco com Email info tradinco com...

Reviews: