TRAQC-20 LPC
TRAQC-20 LPC 0 13-08-2019 Page: 17
The figure shows the flow chart for the AUTO mode. Based on the table, one will know which cycles
are being carried out and the meaning of each.
Phase
Description
1
Wait -Start with OK button-
2
Wait -Can be configured in the MENU-
4
Duration of zeroing the internal sensor
7
Duration of point of zero
9
Time of the adjustment till the next step
10
Stop time
13
Wait (target pressure)
15
Time of the adjustment till the next step
16
Duration at point of zero
18
Stop -Can be configured in the MENU-
Between phases 10 and 16, the calibrator data can be evaluated and recorded if a stable value is
displayed. The AUTO mode can be exited by pressing the MENU button. VENT will be shown In the
top left of the display.
5.4 Leakage (LEAK)
In the CTRL mode you it is possible to check connected pressure device for leaks.
Settings
To check pressure devices for leaks, you have to work in the CTRL mode.
·
Enter the desired pressure with the buttons UP/ DOWN or the buttons 0%/ 100%.
·
As soon as the required pressure is reached and the value is stable push the LEAK button. The
internal pump will be switched off and pressure is held.