TRAQC-20 LPC 0 13-08-2019 Page: 16
5.3 Operating mode 'AUTO'
Store specific profiles in the AUTO mode. Use the AUTO mode to calibrate different pressure
devices with identical values. The defined profile can be repeated a several times.
Settings
·
Push the button MENU. Confirm MODUS with OK and push UP/ DOWN till AUTO is displayed.
Confirm with OK.
·
The submenus RANGE, UNIT and STEPS aren‘t relevant in AUTO mode.
·
Push the buttons UP/ DOWN till SETTINGS is displayed and confirm with OK. Submenu ZERO
isn‘t relevant in this mode.
·
Push the buttons UP/ Down till AUTO-MODUS is displayed and confirm with OK.
o
RANGE: Relevant pressure range
o
UNIT: Pressure unit (see settings)
o
STEPS: The pressure range is divided into sections.
Further settings of AUTO mode can be set in the submenu CONFIGURATION.
·
Cycles: Set the number of cycles that have to be realized in the AUTO mode.
·
T Start: Wait for vent of the internal sensor.
·
Stop: Wait after reaching the maximum pressure.
·
T Pause: Wait between two cycles
·
AUTO-ZERO: Automatic zeroing at the end of every cycle.
Exit of the sub menu by pushing the button MENU.
Use
After exit of the MENU, the calibrator is on wait status. All information about AUTO mode are shown
on the display.
1) Set pressure
2) Pressure unit
3) Phase 1: it has been vented (pressure = 0)
4) Percentage display of the current set pressure
5) First of ten cycles
6) Operating mode