Operational instructions

The table below shows a general description of the available modes. Please see the sections below
for more details.

Operating mode

Application

Function

CTRL (control)

Calibration of sensors and other
devices

Pressure is built up and maintained.
Target pressure may be percentage
or absolute

MESS (measure)

Measurement of differential and
relative pressure

The calibrator analyzes the
pressures that are connected to the
devices

AUTO (automatic)

Stored pressure profiles are running Pumps build up the pressure, based

on the stored pressure profiles

MENU

Customer specific adaption to the
respective application

Setup function for different operating
modes

5.1 Operating mode 'Control' (CTRL)

The CTRL mode is used for the calibration of pressure sensors and other pressure instruments. In
CTRL mode, the pump is active and regulates according to the desired pressure. The internal
sensor provides the current value on the display.

Setting

1. Press the MENU button. Confirm the menu point MODE with OK and press the UP / DOWN

button until CTRL appears. Confirm with OK

2. In the submenu RANGE the upper limits of the pressure range are defined (UP / DOWN and

LEFT / RIGHT buttons) and confirmed with OK

3. In the submenu UNIT the pressure units are defined (UP / DOWN buttons) and confirmed with

OK

4. In the submenu STEP the step sequence in % is defined (UP / DOWN buttons and LEFT /

RIGHT buttons), confirmed with OK

5. With the MENU button, you can exit the submenu

Use

The unit is located at exit from the menu in VENT mode, ie the sensor of the calibrator is vented. On
the middle of the top, the target pressure is displayed. On the bottom of the display, the per cent
value is shown. When leaving the menu, this value is 0%. Switch between target pressure and per
cent value with the OK button.

·

The percentage value can be changed with the UP / DOWN (in the fixed step sizes) and using the
0% / 100%.

·

The target pressure can only be changed with the UP / DOWN. With 0% / 100% to the
appropriate place, the target pressure be jumped.

Example

Target pressure: By pressing the OK button, you can switch between the percentage and target
pressure. value. Press the OK button until the value (top center display) outlined in bold. Press the
0% or 100% button to jump to the selected location. Choose the required value with the UP and
DOWN button. Push the OK button to exit target pressure. The marker moves to the percentage
value. (middle of the display)

When the calibratos displays a stable value, the data can be recorded:

·

Nominal value: Right below

·

Unit: Right above

·

Actual value: On the display of the connected device

·

Reference value: Middle of the display

Summary of Contents for TRAQC-20 LPC

Page 1: ...TRAQC 20 LPC Version 0 Date 13 08 2019 User manual...

Page 2: ......

Page 3: ...ctions on the rear 11 2 Voltage supply 11 3 Front display of the Traqc 20 LPC 12 4 Pressure connections 12 5 Overpressure protection 13 5 Operational instructions 14 1 Operating mode Control CTRL 14 2...

Page 4: ...harge for parts or labour or will provide a replacement in exchange for the defective product In order to obtain service under this warranty Customer must notify TRADINCO INSTRUMENTS of the defect bef...

Page 5: ...ivery Traqc 20 LPC Power supply unit Operating instructions Calibration certificate per DIN EN 10204 optional DKD DAkkS certificate Cross check scope of delivery with delivery note 1 3 Symbols used Wa...

Page 6: ...tructions 2 1 Intended use The Traqc 20 LPC is used for testing and calibrating pressure sensors This instrument is not permitted to be used in hazardous areas The instrument has been designed and bui...

Page 7: ...ed personnel who have the qualifications described below Skilled personnel Skilled personnel are understood to be personnel who based on their technical training knowledge of measurement and control t...

Page 8: ...balance Automatic in adjustable time intervals Manual ZERO button Housing Instruments version desktop instrument with carry handle Dimensions 102 6 x 257 x 271 mm without carry handle Weight 4 5 kg In...

Page 9: ...n test volume Generation volume 1 2 l min Communication Interface RS 232 and USB Response time 1 value s Approvals and certificates CE conformity EMC directive 2014 30 EU EN 61326 1 2013 EN 61000 3 2...

Page 10: ...TRAQC 20 LPC TRAQC 20 LPC 0 13 08 2019 Page 10...

Page 11: ...he output voltage can result in life threatening voltages at the instrument Only the original power supply unit included in initial delivery Mascot model 9926 should be used Design the wiring particul...

Page 12: ...In MENU mode to go into the various sub menus 7 0 button In CTRL mode you can use the 0 button to change the value from 100 to 0 In sub menu for setting the decimal 8 Zero button Button for manual ze...

Page 13: ...er is made via a 6 6 x 11 mm pressure connection The hose diameter should be 6 mm 4 5 Overpressure protection The maximum permissible pressure must not be exceeded Pressure range Overload 1 mbar 5 tim...

Page 14: ...and confirmed with OK 3 In the submenu UNIT the pressure units are defined UP DOWN buttons and confirmed with OK 4 In the submenu STEP the step sequence in is defined UP DOWN buttons and LEFT RIGHT b...

Page 15: ...ubmenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu...

Page 16: ...and confirm with OK o RANGE Relevant pressure range o UNIT Pressure unit see settings o STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIG...

Page 17: ...ration at point of zero 18 Stop Can be configured in the MENU Between phases 10 and 16 the calibrator data can be evaluated and recorded if a stable value is displayed The AUTO mode can be exited by p...

Page 18: ...T Selection of the pressure unit Pa hPa kPa mbar bar Torr mmHG inHG psi mmH20 inH20 STEPS Setting of the step size in percent MODE Selection of the operating mode CTRL control Calibration of pressure...

Page 19: ...e control via USB is not different from the control via RS232 6 2 RS 232 interface To connect the RxD TxD and GND lines are required The connection is made with an in line serial cable 1 1 male female...

Page 20: ...ld time phase 10 and 16 1 10000 time in seconds atp 1 10000 Pause time phase 18 1 10000 time in seconds atr 1 10000 Tolerance band 1 10000 tolerance in 0 01 FS full scale of the instrument measuring r...

Page 21: ...s smm a c m Define the operating modes of the instrument a start automatic mode c start control mode m start measure mode ssw 1 100 Step size 1 100 step size from operating the up and down buttons in...

Page 22: ...CTUAL pressure Example 0 05 pk CR Reading current unit as text Example Mbar 6 5 Driver The USB interface driver is from FTDI and can be download on their website for all major operating systems Window...

Page 23: ...PC 0 13 08 2019 Page 23 7 Accessories Interface cable RS 232 interface cable USB interface cable Voltage supply Power supply unit Certificates DKD DAkkS certified accuracy Miscellaneous Robust transpo...

Page 24: ...Tradinco Instrumenten Apparaten B V Randonstraat 250 2718 TB Zoetermeer The Netherlands info tradinco com www tradinco com www tradinco com...

Reviews: