12.
Troubleshooting
Problems
Possible causes
Remedies
The sensor does not light up
Batteries empty.
Replace the batteries.
Fault in circuitry.
Contact Tractel®
No return to initial zero
Tare function enabled.
Disable the Tare function and
display the “GROSS” value of
the force.
Sensor permanently
deformed due to
incorrect handling,
excessive overload or
compression.
The product must not be used
before it is verified by Tractel®.
No change in the display or inconsistent display.
Malfunctioning sensor or
circuitry.
Reset: Switch the sensor off and
then on.
If the malfunction persists, contact
Tractel®
Linearity or accuracy problem.
Malfunctioning sensor or
circuitry.
Contact Tractel®
Sensor subject to
compressive or torsional
stress.
Eliminate compressive or torsional
stress on the product.
If the malfunction persists, contact
Tractel®
13.
Declaration of conformity
Copy of the declaration of conformity in the appendix