3.3.7 Resetting the timer (Fig 1.5)
In this mode pressing the middle switch will display the following screen which will flash rapidly to
warn the user this will reset the timer and the elapsed time data will be lost:
The time remaining is displayed in the top left corner. This can be reset to the programmed
default cycle time by pressing the middle switch. After the switch press it will jump to the screen
shown in Fig 1.1. Should this mode be entered by mistake do not press the middle button. Turn
off the incubator and then turn back on which will exit this mode and return to the menus without
resetting the timer.
3.3.8 Incubation cycle time setting (Fig 1.6)
In this mode the default cycle time can be changed from 24 hours to anywhere between 0 and 24
hours.
Before adjusting the cycle time reset the timer as described in section 3.3.7.
To adjust the time press the middle switch. The screen will change to:
Pressing the left switch will decrease the time and pressing the right switch will increase the time.
When the desired time has been set press and hold the middle switch to save the settings. The
display will return to the one shown in Fig 1.1.
20.9°C
20.9°C
New <SEL
OFF
15:45 TIMER A
RST
20.9°C
20.9°C
Tim
<ADJ
OFF
18:00
TIMER A
-
Save +
Summary of Contents for AquaSafe WSL50
Page 1: ...AquaSafe WSL50 Water Safety Laboratory Instruction Manual...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual...
Page 28: ...Section 4 AquaSafe Membrane Filtration Instruction Manual Contents...
Page 44: ...Section 5 AquaSafe Turbidity Tube Instruction Manual...
Page 47: ...Section 6 AquaSafe Block Tester Instruction Manual for WSL Range...