background image

Head office

TR Equipment AB, Sweden

Phone: +46 140 38 50 60

[email protected]

www.trequipment.com

Group Subsidiaries

TR Group Inc, US, Phone: +1 (800) 752 6900 

TR Equipment Ltd, UK, Phone: +44 (0) 844 335 8386

TRanspoRehaMed GmbH, Germany, Phone: +49 (0) 212 645850-0

Hygiene Equipment 

Designed for People

TR 3200

Shower trolley

Hygiene Equipment 

Designed for People

TR reserves the right to make 

design changes and/or  

improvements without notice.

Registered at Medical Products 

Agency, Sweden, 

as a MD Class 1 product, 

MDD 93/42EEC.

2018-01-01

Service manual

Always study this manual carefully before operating the 
Shower Trolley.

Operating instructions and recommended maintenance  
procedures are explained in the following pages.

If you have any questions regarding the operation or  

maintenance of your Shower Trolley and cannot find the an

-

swer in this manual, please contact us.

Only qualified staff who are trained and certified and who 

can understand all operating and safety procedures may 

operate and service this equipment.

The Trolley is designed for years of trouble free operation, with 
proper use and maintenance.

Summary of Contents for TR 3200

Page 1: ...l Products Agency Sweden as a MD Class 1 product MDD 93 42EEC 2018 01 01 Service manual Always study this manual carefully before operating the Shower Trolley Operating instructions and recommended ma...

Page 2: ...to replace the control box 18 How to replace the batteries 20 How to replace the circuit board 22 How to replace the magnetic charger connector and cover 24 How to replace the fuse and fuse cabling 2...

Page 3: ...Hygiene Equipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 3 Spare parts list TR 3200...

Page 4: ...Hygiene Equipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 4 3 1 4 5 6 10 11 9 13 14 15 16 17 18 12 2 Spare parts TR 3200 TR 3200 L onboard batteries 7 8...

Page 5: ...iene Equipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 5 Spare parts TR 3200 TR 3200 L onboard batteries 29 27 22 21 28 20 24 25 26 23 33 37 32 31 36 35 34 19...

Page 6: ...10 7333474 PVC plate 1 for model L 1262 780mm 10 7333474 02 PVC plate 2 for model L 872 780mm 11 851895 02 Lifting column 12 3233204 Chassi 13 851899 Hand control 14 851760 Castor with brake with Grey...

Page 7: ...Description 28 3430021 Mounting plate 29 851897 02 Control box 851897 02 30 3333563 Metal Lid 31 819148 Charger inlet connection 32 819150 Charger inlet cover 33 0541226 Battery charger 110 240V Incl...

Page 8: ...Hygiene Equipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 8 Spare parts TR 3200 R TR 3200 LR removable batteries 17 18 15 16 14 9 11 10 1 12 13 6 5 2 4 3 7 8...

Page 9: ...Hygiene Equipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 9 Spare parts TR 3200 R TR 3200 LR removable batteries 19 20 21 24 23 25 26 22...

Page 10: ...10 7333474 PVC plate 1 for model L 1262 780mm 10 7333474 02 PVC plate 2 for model L 872 780mm 11 851895 02 Lifting column 12 3233204 Chassi 13 851899 Hand control 14 851760 Castor with brake with Gre...

Page 11: ...ll protected from possible damages 3 If necessary carefully clean the threads from Loctite with an M10 tap 4 Insert the bolt and washer through the castor body 6 Fasten the bolt 7 Assemble the cap by...

Page 12: ...an M10 tap 7 Insert the bolt and washer through the castor body 9 Fasten the bolt 4 Place the straight steering plate onto the chassis Note the position of the angle on the plate and the pins 5 Adjus...

Page 13: ...cable cables and remove the stretcher 4 Remove the four M8x30 bolts attached to the lifting column and remove the bracket with the dampers 5 Turn the chassi around and remove the four M8x30D bolts th...

Page 14: ...s 9 Put the stretcher back onto the chassis 11 Connect the actuator cable with the cable from the controlbox 12 Fasten the cables with plastic stripes again and the put plastic plate and mattress back...

Page 15: ...plastic plate 1 Unscrew the four M10x12 bolts holding the stretcher 3 Disconnect the actuator cables 2 Cut off the plastic stripes to loosen the cables connected to the actuator 4 Unscrew the lid for...

Page 16: ...tcher position and lock castor brakes and straight steering castors Remove mattress and plastic plate Note When replacing the stretcher we recommend to also replace bolt and screws If the bolts and sc...

Page 17: ...teering castors Remove the mattress Tighten the M6x20 screw and M6 locking nut 50 only It must be possible to raise and lower the side rails 3 Insert screw and nut according to the picture Screw Nut S...

Page 18: ...the power The emergency stop shall be activated until the new control box is installed 5 Disconnect the lifting column and hand control cables and cut the cable ties 1 Unscrew the four M5x12D screws...

Page 19: ...h the control box to the right How to replace the control box 7 Pull out the cable from the enclosure Service instructions Onboard batteries 8 Connect the control box according to the wiring diagram 9...

Page 20: ...s and plastic plate Press the Emergency stop button to cut the power to the circuit board The button must be pressed until the batteries are replaced and plugged in 2 Unscrew the plastic lid on the ba...

Page 21: ...elease the Emergency stop by turning it clockwise a click shall be heard from the control box 9 Put the plastic lid back onto the plastic enclosure Fasten it by tightening the fours screws on the lid...

Page 22: ...board See wiring diagram on page 27 5 Insert the new circuit board into the battery box 6 Release the emergency stop by turning it clockwise a click shall be heard from the control box Service instru...

Page 23: ...for People 23 7 Put the plastic lid back onto the plastic enclosure Fasten it by tightening the four screws on the lid 8 Put the metal lid back onto the stretcher and fasten it by tightening the four...

Page 24: ...box by loosening the four screws on the plastic lid 3 Remove the fuse from the fuse holder to cut the power 4 Remove the contact block from the emergency stop by squeezing the clip 6 Loosen the connec...

Page 25: ...r and cover into the battery box 8 Fasten the connector by turn ing the fastening nut clockwise Use a pliers for proper fastening 11 Insert the fuse into the fuse holder A click shall be heard from th...

Page 26: ...The button must be pressed until the fuse cabling is mounted 4 To replace the fuse cabling remove and lead from the batteries Insert the new fuse ca bling incl the fuse by connecting and leads to the...

Page 27: ...uipment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 27 Charger MagCode connector Wiring diagram Service instructions Onboard batteries Red Brown 1 2 Yellow Green C...

Page 28: ...the battery from the control box connector 5 Loosen and remove handcon trol and lifting column cables 3 Cut the cable ties holding the control box cable 6 Press the lever and push the control box to t...

Page 29: ...connector How to replace the battery Service instructions removable batteries 2 Loosen and remove the battery 3 Attach the new battery onto the bracket Make sure the bat tery is properly attached 4 C...

Page 30: ...mattress and plas tic plate 2 Cut the cable ties holding the hand contol cable 3 Remove the bracket on the control box 4 Attach the new hand control 5 Fasten the bracket on the control box 6 Fasten th...

Page 31: ...ment Designed for People TR 3200 Shower trolley Hygiene Equipment Designed for People 31 Control box connection guide Service instructions removable batteries 1 Hand control 2 Lifting column 3 Lifting...

Page 32: ...started up Initialization Calibration Use the hand control and press both the UP and DOWN button simultaneously Keep pressing both buttons through the complete Initialization calibration After approxi...

Page 33: ...atteries are charged and works properly Make sure that it s possible to raise lower and tilt the stretcher with the hand control Make sure that it s possible to raise and lower the side rails Make sur...

Page 34: ...are fully functioning Make sure that raise and lowering function works properly the entire lift range Make sure the tilting functions works properly Annual maintenance It s recommended to consult an a...

Page 35: ...d dried Yes No Comment Action Chassis Are screws and fasteners tightened Yes No Comment Action Chassis Any signs of wear or damages Yes No Comment Action Mattress and pillow Are mattress and pillow cl...

Page 36: ...e the hand control Yes No Comment Action Lifting columns Is it possible to tilt the stretcher Use the hand control Yes No Comment Action Lifting columns Calibrate Initialize the lifting col umns See i...

Page 37: ...rol box Yes No Comment Action Castors Replace castors Yes No Comment Action Shower trolley Approved and OK to use Yes No Comment Action Note The shower trolley should be immediately taken out of opera...

Page 38: ...ohol content Cleaning process Lightly brush off loose dirt particles first then rinse material with clear water Spray on the cleaning liquid detergent in recommended concentration or apply with a clea...

Reviews: