Page 16
5. pH Calibration
5.1 Calibration Procedure
1. Plug the pH sensor into the
pH
socket. Temperature measurements
are made via the Conductivity sensor, so this needs to be connected
for Automatic Temperature Compensation.
2. Switch the meter on.
3. Ensure that temperature has been calibrated or manually set (see
sections 7.1 & 7.4). If the decimal point in the temperature reading is
shown by a “
“, then the temperature value is not calibrated.
4. Remove the wetting cap from the pH sensor. Rinse the pH and
Conductivity sensors in distilled water and blot them dry.
5. Ensure that you are using buffers which have been selected for
automatic buffer recognition. See section 17 for a detailed
explanation.
6. Place both sensors into a small sample of pH7.00 (or pH6.86) buffer,
so that the bulb and reference junction are both covered, as per the
diagram below.
DO NOT
place the sensors directly into the buffer
bottle.