background image

 

Section 4 – Peristaltic Dosing Pumps

 

 

14 

To fit a new Pump Tube… 

 

Turn the roller holder so 
that the rollers are in a 
horizontal line. Push the 
left hand side 
compression fitting into 
its seat, round side 
facing in. Begin turning 
the roller holder 
clockwise, so that the 
roller holder acts as a 
lever which pulls the 
pump tube in as it turns. 
After approximately 3/4 
of a turn, the right hand 
side compression fitting 
can be pushed in, round 
facing in. 

The new pump tube has 
now been fitted.

 

 

 

 

Replace the see-through 
protector plate on the 
front of the pump. This 
is essential to protect 
the rollers and tube from 
dirt and moisture.

 

Re-fit the Input and 
Output tubing to the 
compression fittings. 
See the procedure 
detailed in section 2.4.

 

The power to the 

easy

FLOW-VS

 

can 

now be re-connected 
and switched on.

 

Warning :

  Use only a genuine replacement Tube supplied by TPS. Use of other tubing materials, 

sizes  or  lengths  may  place  undue  stress  on  the  rollers,  causing  them  to  wear 
prematurely. This type of wear is not covered by warranty. 

 

Summary of Contents for easyFLOW-VS

Page 1: ...td ABN 30 009 773 371 Unit 6 253 Leitchs Road Brendale QLD Australia 4500 Phone 07 3205 8027 International 61 7 3205 8027 Fax 07 3808 4871 International 61 7 3808 4871 Email tps tps com au Web www tps...

Page 2: ...ions 9 2 6 External Control Settings 10 2 7 Setting the Pump Speed to 50 11 Section 3 Operating Instructions 12 3 1 Manual Operation 12 3 2 External On Off Control Operation 12 3 3 External 4 20 mA or...

Page 3: ...mation section to ensure that you have received all accessories and any optional items you may have ordered Section 2 Installation This section provides details on installing the easyFLOW VS including...

Page 4: ...control mode using a 4 20mA or 0 1V input 3 External Control Indicator The External Control indicator LED lights when the easyFLOW VS is set for either external on off control or external speed contro...

Page 5: ...al drum Part No 116354 easyFLOW VS Handbook Part No 130050 24V 500mA 1A AC DC Adaptor Part No 130106 Australia 240V 2 Flat Pins 130108 Euro Asia 220V 2 Round Pins 130110 USA 110 120V 2 Flat Pins 13011...

Page 6: ...C 500mA 1 Amp AC DC Adaptor to suit country of destination is supplied The easyFLOW VS unit draws approximately 200 mA of current Standard Control Dose Rate control allows selection of OFF Dosing Rate...

Page 7: ...bottom of the chemical drums to the dosing pumps is 2 metres 3 The power lead from the AC DC Adaptor reaches from a mains power point to the easyFLOW VS without straining The external control cable be...

Page 8: ...provided is not compatible with the container drill a 1 2 or 13mm hole into the lid of the container and fit the compression fitting Drill a second 2mm diameter vent hole to avoid suction build up 2 4...

Page 9: ...lation 9 2 5 Mount the easyFLOW VS and Make Electrical Connections For wall mounting use minimum 3mm or 1 8 stainless steel bolts screws or masonary anchors A drilling template is provided at the back...

Page 10: ...cess the circuit board s settings remove the four screws at the corners of the front panel and remove the lid Please use caution to avoid damaging any connecting cables In the table below Switch under...

Page 11: ...fitting is provided for these users To select 50 speed operation 1 Remove the front panel and locate the jumper settings as per the instructions in section 2 6 2 Set jumper J3 to On to select 50 speed...

Page 12: ...W VS is under external control 3 It is not necessary to set the Dose Rate control as the pump speed is determined by the 4 20 mA or 0 1 V signal 3 4 External 4 20 mA or 0 1 V Control Operation with Ma...

Page 13: ...he Pump Tubing may be replaced every year as a precautionary measure To remove the old Pump Tube 1 IMPORTANT Place the pump suction tubing into a bucket of fresh water Prime the pump with water by swi...

Page 14: ...an be pushed in round facing in The new pump tube has now been fitted 2 3 4 Replace the see through protector plate on the front of the pump This is essential to protect the rollers and tube from dirt...

Page 15: ...ou feel that this equipment is in need of repair please re read the manual Sometimes instruments are received for repair in perfect working order This can occur where the sensor s simply requires clea...

Page 16: ...Section 5 Warranty 16 This page has been intentionally left blank to ensure that the drilling template is on its own removable page...

Page 17: ...Section 6 Drilling Template 17 Section 6 Drilling Template...

Reviews: