Section 4 – Peristaltic Dosing Pumps

 

 

13

Section 4. –  Peristaltic Dosing Pumps 

4.1 

Lubricating the Pump Tubing 

The 

easy

FLOW-VS

 

is  shipped  with  the  correct  lubrication  on  the  peristaltic  pump  tubing.  This 

lubrication  should  be  re-applied  every  year.  Suitable  Silicon  lubricant  is  available  from  TPS  in  a 
3mL syringe (part no 130103). 

To lubricate the pump tubing… 

1.  Remove the pump tubing, as per the details in section 4.2, below. 

2.  Smear a small quantity of Silicon grease all around the tube. A thin film is all that is required. 

DO  NOT  use  petroleum  based  oil  or  lubricants,  as  these  will  cause  the  Santoprene 
rubber to perish. Use only Silicon based lubricant. 

4.2 

Replacing the Pump Tubing 

Replace the Pump Tubing if it gets worn, begins to leak, or pumping volume is reduced. The Pump 
Tubing may be replaced every year as a precautionary measure. 

To remove the old Pump Tube… 

IMPORTANT 

Place the pump suction 
tubing into a bucket of 
fresh water. Prime the 
pump with water by 
switching on at full 
speed to 3 minutes. This 
ensures that there is no 
dangerous chemical in 
the tube when it is 
disconnected. 

 

 

 

Switch off the power and 
disconnect the power 
cord.

 

Remove the Input and 
Output tubing from the 
compression fittings. 
This is simply a reversal 
of the procedure 
detailed in section 2.4.

 

Remove the see-
through protector plate 
on the front of the pump.

 

 

Turn the roller holder so that the rollers are in a horizontal line. Pull the left 
hand side compression fitting from its seat. Begin turning the roller holder 
clockwise, so that the roller holder acts as a lever which pulls the pump tube 
out as it turns. After approximately 3/4 of a turn, the right hand side 
compression fitting can be pulled out. 

The old pump tube has now been removed. 

 

Summary of Contents for easyFLOW-VS

Page 1: ...td ABN 30 009 773 371 Unit 6 253 Leitchs Road Brendale QLD Australia 4500 Phone 07 3205 8027 International 61 7 3205 8027 Fax 07 3808 4871 International 61 7 3808 4871 Email tps tps com au Web www tps...

Page 2: ...ions 9 2 6 External Control Settings 10 2 7 Setting the Pump Speed to 50 11 Section 3 Operating Instructions 12 3 1 Manual Operation 12 3 2 External On Off Control Operation 12 3 3 External 4 20 mA or...

Page 3: ...mation section to ensure that you have received all accessories and any optional items you may have ordered Section 2 Installation This section provides details on installing the easyFLOW VS including...

Page 4: ...control mode using a 4 20mA or 0 1V input 3 External Control Indicator The External Control indicator LED lights when the easyFLOW VS is set for either external on off control or external speed contro...

Page 5: ...al drum Part No 116354 easyFLOW VS Handbook Part No 130050 24V 500mA 1A AC DC Adaptor Part No 130106 Australia 240V 2 Flat Pins 130108 Euro Asia 220V 2 Round Pins 130110 USA 110 120V 2 Flat Pins 13011...

Page 6: ...C 500mA 1 Amp AC DC Adaptor to suit country of destination is supplied The easyFLOW VS unit draws approximately 200 mA of current Standard Control Dose Rate control allows selection of OFF Dosing Rate...

Page 7: ...bottom of the chemical drums to the dosing pumps is 2 metres 3 The power lead from the AC DC Adaptor reaches from a mains power point to the easyFLOW VS without straining The external control cable be...

Page 8: ...provided is not compatible with the container drill a 1 2 or 13mm hole into the lid of the container and fit the compression fitting Drill a second 2mm diameter vent hole to avoid suction build up 2 4...

Page 9: ...lation 9 2 5 Mount the easyFLOW VS and Make Electrical Connections For wall mounting use minimum 3mm or 1 8 stainless steel bolts screws or masonary anchors A drilling template is provided at the back...

Page 10: ...cess the circuit board s settings remove the four screws at the corners of the front panel and remove the lid Please use caution to avoid damaging any connecting cables In the table below Switch under...

Page 11: ...fitting is provided for these users To select 50 speed operation 1 Remove the front panel and locate the jumper settings as per the instructions in section 2 6 2 Set jumper J3 to On to select 50 speed...

Page 12: ...W VS is under external control 3 It is not necessary to set the Dose Rate control as the pump speed is determined by the 4 20 mA or 0 1 V signal 3 4 External 4 20 mA or 0 1 V Control Operation with Ma...

Page 13: ...he Pump Tubing may be replaced every year as a precautionary measure To remove the old Pump Tube 1 IMPORTANT Place the pump suction tubing into a bucket of fresh water Prime the pump with water by swi...

Page 14: ...an be pushed in round facing in The new pump tube has now been fitted 2 3 4 Replace the see through protector plate on the front of the pump This is essential to protect the rollers and tube from dirt...

Page 15: ...ou feel that this equipment is in need of repair please re read the manual Sometimes instruments are received for repair in perfect working order This can occur where the sensor s simply requires clea...

Page 16: ...Section 5 Warranty 16 This page has been intentionally left blank to ensure that the drilling template is on its own removable page...

Page 17: ...Section 6 Drilling Template 17 Section 6 Drilling Template...

Reviews: