Step by Step Procedures

Note: 

The 623 incorporates a differential manometer.
This means when a single hose is connected to
the (+) port and pressure is applied, the 623 will
read positive pressure. If a single hose is con-
nected to the (-) port and pressure is applied,
the 623 will read a negative pressure. If both
ports are connected to a pressure, the 623 will
read the difference (P1 - P2) between the two.

When performing single pressure measurments
(non-differential) use the positive port and leave
the negative port open.

8

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for 623

Page 1: ...623 Digital Manometer Instruction Manual www com information itm com 1 800 561 8187 ...

Page 2: ...5 Technical Data International Symbols Page 6 Features Benefits Page 7 Pressure Conversion Table Page 8 Step by Step Procedures Page 10 Accessories Specifications Page 11 Maintenance Page 12 Trouble shooting guide Page 13 EC declaration of conformity www com information itm com 1 800 561 8187 ...

Page 3: ...ember to return your product warranty registration card Product Description The TPI 623 is a hand held pressure meter with supe rior resolution Auto Range 100 0 to 100 0 mbar 40 0 to 40 0 inH2O Resolution 0 001 inH2O 0 01 inH2O Accuracy 2 inH2O 0 02 inH2O 2 inH2O 1 of reading Over Pressure Protected to 50 inH2O Power Supply 9V 3 www com information itm com 1 800 561 8187 ...

Page 4: ...at the connecting hose you are using is free from kinks or splits Double check all connections before testing NEVER Connect the Instrument to an unknown pres sure source if the source is twice the working pressure of the instrument the sensor will be damaged and may cause injury to the user Tamper with the instrument or attempt to modify it in any way Otherwise accident and injury may occur and th...

Page 5: ... dispaly ZERO will temporari ly display in the lower left of the LCD Smoothing function to stabilize rapidly changing pressure readings UNITS Selects 7 units of pressure kPa PSI inHg mmHg inH2O mmH2O mbar Turns the back light on automatically turns off after 30 seconds 5 www com information itm com 1 800 561 8187 ...

Page 6: ...ize erratic readings It displays the average value of 4 values previously measured Zero Function Provides immediate push button zeroing for Speed and accuracy Units Change the display value in any one of the following engineering units kPa PSI inHg mmHg inH20 mmH20 and mbar Back Light Provides easy reading of LCD in low light areas 6 www com information itm com 1 800 561 8187 ...

Page 7: ...x 10 2 2 89567 x 10 3 2 953 x 10 4 inH2O 406 8 401 48 13 60 1 27 68 0 5354 3 937 x 10 2 4 014 x 10 3 PSI 14 6955 14 504 0 4912 3 6127 x 10 2 1 1 934 x 10 2 1 422323 x 10 3 1 4503 x 10 3 mmHg 760 750 06 25 400 1 868 51 715 1 7 35433 x 10 2 7 502 x 10 3 mmH2O 10 3327 x 103 10 1976 x 103 3 4544 x 102 25 400 703 07 13 599 1 101972 Pascals 1 01325 x 105 1 000 x 105 3 386 x 103 2 491 x 102 6 8948 x 103 ...

Page 8: ...itive pressure If a single hose is con nected to the port and pressure is applied the 623 will read a negative pressure If both ports are connected to a pressure the 623 will read the difference P1 P2 between the two When performing single pressure measurments non differential use the positive port and leave the negative port open 8 www com information itm com 1 800 561 8187 ...

Page 9: ...o be tested using appropriate hose connection in accordance with the manufacturers instructions 4 The pressure reading will appear on the display To activate the backlight in poor light conditions press the key at any time 5 If the reading is erratic or changing activate the smoothing function by pressing the key 6 To convert the displayed pressure reading to another engineering unit depress the U...

Page 10: ...ifications Auto Range 100 0 to 100 0 mbar 40 0 to 40 0 inH2O Resolution 0 001 inH2O 0 01 inH2O Accuracy 2 inH2O 0 02 inH2O 2 inH2O 1 of reading Over Pressure Protected to 50 inH2O Power Supply 9V Relative Humidity 0 to 80 Operating Temp 0ºC to 40ºC Storage Temp 10ºC to 50ºC Auto Power Off 20 Minutes Input Type Differential 10 www com information itm com 1 800 561 8187 ...

Page 11: ...trument 3 Remove the back cover 4 Remove old battery 5 Replace with new battery 9V NEDA 1604 or equivalent 6 Reassemble the instrument in reverse order from above being sure not to trap any wires between the two halves of the case 7 Re fit the six cross head screws being sure not to over tighten Cleaning Use a mild detergent and a slightly damp cloth to clean the surfaces of the instrument 11 www ...

Page 12: ...ero button has not been pushed Unit still does not zero If reading is greater than 0 05 and constantly fluctuates return to local Service Centre Bat appears on display Replace battery immediately 9V NEDA 1505 or equivalent 12 www com information itm com 1 800 561 8187 ...

Page 13: ... of the member states relating to Electromagnetic compati bility by application of the following standards EN50081 1 1992 Emissions Standard EN50082 1 1992 Immunity Standard To ensure conformity with these standards this instrument must be operated in accordance with the instructions and specifications given in this manual 13 www com information itm com 1 800 561 8187 ...