background image

F. MAINTENANCE

1.

Clean the surface of the instrument and optical lens with a damp cloth.

2.

Replace the 9 Volt battery when the low battery indicator appears on
the LCD.

G. TROUBLE SHOOTING

Symptom

Probable Cause

Temperature seems inaccurate.

• Dirty optical lens.
• Low battery.
• Instrument held too far away 

from target increasing desired 
target area. 

• Target surface has poor 

emissivity qualities.

• Laser sight out of alignment.

371/373 does not turn on.

• Dead or low battery.
• Defective POWER ON switch.
• Broken batter connector.

Laser does not turn on.

• Low battery.
• Broken connector to Laser.

Test Products International, Inc.

9615 SW Allen Blvd., Ste. 104

Beaverton, OR USA  97005

503-520-9197 • Fax: 503-520-1225

[email protected]

Test Products International, Ltd.

342 Bronte Road South, Unit 9

Milton, Ontario  L9T5B7

Canada

905-693-8558  •  Fax: 905-693-0888

L373M • 4/24/00  copyright © 2000 Test Products International, Inc.

Infrared Thermometers 375,376,377

from 

Test Products International, Inc.

The Value Leader

TM

Visit www.tpi-thevalueleader.com  

Common Emissivity Values

Material

Emissivity

Aluminum (oxidized)

0.25

Aluminum (polished)

0.1

Asbestos

0.95

Brass (oxidized)

0.6

Brass (polished)

0.1

Carbon

0.75

Carborundum

0.85

Cardboard

0.9

Cast Iron (polished)

0.2

Cast Iron (rusted)

0.95

Chromium (polished)

0.1

Concrete

0.7

Copper (oxidized)

0.8

Copper (polished)

0.05

Gold (polished)

0.1

Iron plate

0.7 to 0.85

Lead (oxidized)

0.3

Lead (pure)

0.1

Marble

0.9

Matte black paint 

0.95

Nickel (pure)

0.1

Nickel plate (oxidized)

0.95

Paper

0.9

Plaster

0.9

Plastics

0.8 to 0.95

Quartz

0.9

Red Brick

0.75 to 0.9

Rubber (rough)

0.98

Rubber (smooth)

0.9

Silica

0.4

Silver (polished)

0.1

Stainless Steel (other)

0.2 to 0.6

Stainless Steel (polished)

0.1

Steel (ground sheet)

0.6

Steel (Mild)

0.3 to 0.5

Steel plate

0.9

Timber

0.8 to 0.9

Water

0.98

Zinc (oxidized)

0.1

Reviews: