TIC-LF620E

 

Introduction 

Combined with a multi-functional converter LF620 
(combined type) or LF622 (separate type) equipped 
with its original noise-suppression circuit and 
advanced algorithms. IR (Infrared) switches enable 
parameter setting of the converter without removing 
the cover. Flow direction can be set in either way, and 
its 128 x 128 dot matrix LCD display allows the LCD 
to be rotated electronically to 90, 180 and 270 degrees 
without opening the cover. The terminal block in LCD 
side make easy to wire in case of the combined type. 

 

*1: HART protocol (Highway Addressable Remote Transducer) is a 

communication protocol for industrial sensors recommended by the 

HCF (HART Communication Foundation). 

       

 ** DevComm2000 Smart Device Communicator available 

through TIC for performing HART device configurations on 
PC or laptop . 

*2: PROFIBUS is the communication protocol for factory automation and 

process automation that the PROFIBUS Organization recommends. 

Instead of analog control with a conventional analog signal (4-20mA), 

it is one kind of the fieldbus which digitizes all signals. Flowmeters 

support PROFIBUS-PA. 

*3: Modbus is the communication protocol that Modicon Inc. developed. 

Physical layer is RS485. 

 

 

  

 

LF620  

LF622 

                    LF620F                    LF622F 

Figure1.  LF620 series Flowmeter Converters 

 

                      
 
 
 

Specifications 

 

Model LF620 and LF622 converters 

Input signals 

Analog signal

 —

 

the voltage signal from detector, 

proportional to process flow rate (For LF622 
separate type converter).

 

Digital input DI 

Signal type: 20 to 30Vdc voltage signal 
Input resistance: 2.7k

 

Number of inputs: one point 

Note:

 DI cannot be used with the Modbus 

communication. 

DI function 

— One of the following functions can 

be assigned to the optional DI signal. 

Range switching 

— Selects either the higher or 

lower range in the unidirectional or 
bidirectional 2-range setting.

 

Totalizer control 

— Starts and stops the built-in 

totalizer. 

Fixed-value outputs 

— Outputs fixed-values for 

current and pulse outputs. 

Zero adjustment 

— Executes zero adjustment 

(on-stream at zero flow rate). 

Output signals 

Current output: 

4–20mAdc (load resistance  0 to 750

Note: 

The current output cannot be used with 

the PROFIBUS-PA communication. 

Digital outputs

 —  Two points are available as 

follows. 

Digital output DO1 : 

Output type: Transistor open collector 
Number of outputs: One point 
Output capacity: 30Vdc, 200mA maximum 

Note:

 DO1 cannot be used if Modbus 

communication connection is 3 lines. 

Digital output DO2 : 

Output type: Solidstate relay output (non 
polarity) 
Number of outputs: One point 
Output capacity: 150Vdc, 150mA maximum 
or 150 V ac (peak to peak), 100mA maximum 

Note:

 DO2 cannot be used with the Modbus 

communication. 

 

 

 

Field Intelligent Device Series 

Electromagnetic Flowmeter Converter 

 

LF620,LF622

Certification number

 

Z01207

 

 

 

 

 

M.A. Selmon Company, Inc
4 Oxford Rd.
Milford, CT 06460

203-377-3525

Summary of Contents for LF620

Page 1: ...d LF622 converters Input signals Analog signal the voltage signal from detector proportional to process flow rate For LF622 separate type converter Digital input DI Signal type 20 to 30Vdc voltage signal Input resistance 2 7k Number of inputs one point Note DI cannot be used with the Modbus communication DI function One of the following functions can be assigned to the optional DI signal Range swi...

Page 2: ...evice Max cable length 1 2km Note Note This length is specification of 3 line connection LCD display Full dot matrix 128 128 dot LCD display back light provided A parameter change will rotate the display Parameter settings Parameters can be set as follows IR Switches Three key switches are provided to set configuration parameters Digital communication HART PROFIBUS or Modbus is needed to set param...

Page 3: ...10 to 150Hz with acceleration of 9 8m s2 Vibration of 30Hz with 29 4 m s2 in 4h in each direction will not cause any defect to unit Note Avoid using the flowmeter in an environment with constant vibration Converter LF622 dimensions and weights See Figure 2 for Separate type MTBF 220 000 hours at 25 deg C 77 deg F based on MIL HDBK 217F ...

Page 4: ... and LF622F Note Cable glands are not provided for LF622F cFMus approved type Refer to the part Cable connection port at detector LCD display Attachment I O cable ground Excitation cable ground Power supply cable ground Option cable ground Signal cable ground IR Switch Plate ...

Page 5: ...th 100Ω or less ground resistance or Modbus Power switch External double pole power switch NOTE In thecase of the integral type the signal line A B G andthe excitation line E X Y are the connector joints Grounding with 100Ω or less ground resistance 1 Locate an external double pole power switch on the power line near the flowmeter within easy reach of operation Use the appropriate switch rating as...

Page 6: ...lled steel conduit Power switch External double pole power switch Symbol Description Cable L1 Power supply Power cable L2 GND Ground for arrester FG Frame ground DI Digital Input 20 30Vdc I O cable DO1 Digital Output 1 DO2 Digital Output 2 COM Signal Common for DI DO1 DO2 Current Output 4 20mAdc or PROFIBUS Shielded cable for PROFIBUS PA X Excitation cable for LF622 LF622F only Y Excitation Output...

Page 7: ...ion make sure to use a shielded cable to improve noise resistance Furthermore installation of signal cable in metal conduit is recommended 3 General cables are designed for indoor use where cables are not exposed to humidity rain etc When you install cables make sure to check the operating conditions such as the operating temperature range of the cable by contacting its manufacturer 4 When you car...

Page 8: ...ote 1 When PROFIBUS communication is provided current output 4 20mA and HART communication cannot be used When Modbus communication is provided digital output points 1 DO1 and digital output points DO2 digital input point 1 DI HART communication cannot be used Refer Table 2 for more details Note2 Select 110Vdc for test report inspected under the condition of 110Vdc Table2 Communication functions a...