Torrey EchoTherm HP60 SERIES Operating Manual Download Page 1

 
 
 
 
 

OPERATING MANUAL 

 

EchoTherm™ 

PROGRAMMABLE HOT PLATES AND  

STIRRING HOT PLATES 

 

HP60 AND HS60 SERIES 

 
 
 
 

DOCUMENT NUMBER HP60/HS60-1 

Revised 17 January 2011 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

TORREY PINES SCIENTIFIC, INC. 

2713 Loker Ave. West 

Carlsbad, CA 92010 

 

TELEPHONE: (760)-930-9400 

Toll Free: 866-573-9104 

FAX: (760)-930-9480 

E-MAIL: [email protected]

 

WEB SITE: www.torreypinesscientific.com

TABLE OF CONTENTS 

 
PAGE 

Summary of Contents for EchoTherm HP60 SERIES

Page 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Page 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Page 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Page 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Page 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Page 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Page 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Page 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Page 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Page 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Page 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Page 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Page 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Page 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Page 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Page 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Page 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Page 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...