ToronTek G64 User Manual Download Page 1

Instructions to User 

This Manual is written and compiled in accordance with the

 

council directive MDD93/42/EEC for medical devices 

and harmonized standards. The Manual is written for the current Pulse Oximeter. In case  of modifications and 

software upgrades, the information contained in this document is subject to change without notice. 

The Manual describes, in accordance with the Pulse Oximeter's features and requirements, the  main structure, 

functions, specifications,  correct methods for transportation, installation, usage, operation, repair, maintenance and 

storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters 

for details. 

Please read the Manual carefully before using this equipment. These instructions describe the operating procedures to 

be followed strictly. Failure to follow these instructions can cause measuring abnormality, equipment damage and 

personal injury. The manufacturer is NOT responsible for the safety, reliability and performance issues or  any 

measuring  abnormalities, personal injury and equipment damage due to user's negligence of the operation 

instructions. The manufacturer's warranty service does not cover such faults.  

Regrettably, owing to the forthcoming renovation, the specific products you received may not be totally in accordance 

with the description of this User Manual.  

This product is a medical device, and can be used repeatedly. Its useful lifespan is 3 years.  

WARNING: 

Discomfort  or pain can occur if the device is used ceaselessly, especially for microcirculation barrier 

patients. It is recommended that the sensor should not be applied to the same finger for over 2 hours. 

For special patients, there should be a more prudent inspection of the placing process. The device should 

not be clipped on an edema or tender tissue. 

The infrared light emitted from the device, though invisible, is harmful to the eyes. Avoid staring at it. 

The patient cannot use nail enamel or other makeup. 

The patient’s fingernails must be clipped short. 

Please refer to the related content about the clinical restrictions and caution. 

This device is not intended for treatment.  

1 Safety 

1.1 Instructions for Safe Operations 

Inspect the main unit and all accessories periodically to ensure that there is no visible damage that may affect

the patient’s safety, and monitor the performance of cables and transducers. It is recommended that the device 

should be inspected at least once a week. If there is obvious damage, stop using the device. 

Necessary maintenance must be performed by qualified service engineers ONLY. Users are not permitted to

service the device by themselves.

The  Oximeter  must not  be used together with devices not specified  in  the User  Manual.  Only accessories

specified or recommended by the manufacturer can be used with this device.

This product is calibrated before leaving the factory.

1.2 Warning 

Explosive hazard—DO NOT use the Oximeter in an environment with flammable gas, such as some ignitable

anesthetic agents. 

DO NOT use the Oximeter while the patient is measured by MRI and CT. 

People who are allergic to rubber cannot use this device. 

The disposal of this instrument and its accessories and packaging (including batteries, plastic bags, foams and

paper boxes) should follow local laws and regulations. 

Please check the packaging before use to make sure the device and accessories are in complete accordance with

the packaging list or else the device may not work normally.

Please do not measure this device with function test paper for the device’s related information. 

1.3 Attention 

 

Keep the Oximeter away from dust, vibrations, corrosive substances, explosive materials, high temperatures and 

moisture. 

 

Stop operating the Oximeter if it gets wet. 

 

DO NOT use the device immediately after moving it from a cold environment to a warm or humid environment.  

 

DO NOT operate the keys on the front panel with sharp objects. 

 

High temperature and  high pressure steam disinfection of the Oximeter  is not permitted. Refer to the  User 

Manual in the relative chapter for instructions on cleaning and disinfection. 

 

DO NOT immerse the Oximeter in liquid. If it needs cleaning, wipe the surface with medical alcohol using a 

soft material. Do not spray any liquid directly onto the device. 

 

When cleaning the device with water, ensure that the temperature of the water is lower than 60

 

Thin or cold fingers will affect the normal measure of the patient's SpO

2

 and pulse rate. Be sure to clip a thick 

finger such as the thumb or middle finger deeply enough into the probe. 

 

DO NOT use the device on infants or neonatal patients. 

 

This  product is only  suitable  for adults  and  children above four years old    (Weight should be between 15kg-

110kg).  

 

The device may not work for all patients. If you are unable to achieve stable readings, discontinue use. 

 

The update period of data is less than 5 seconds. This is changeable according to different individual pulse rates. 

 

The waveform is normalized. Please read the measured value when the waveform on-screen is uniform and 

steady-going. This measured value is the optimal value and this waveform is the standard one. 

 

If abnormal conditions appear on the screen during the test process, pull out the finger and reinsert it to restore 

normal use. 

 

The device has a lifespan of three years after the first electrified use. 

 

The  wrist strap  attached  to  the  device  is made from non-allergenic  material.  If  you  are sensitive to the wrist 

strap, stop using it. In addition, pay attention to the use of the wrist strap. To avoid  harm, DO NOT wear it 

around the neck. 

 

The device does not have a low-battery alarm function; it only displays the low battery icon. Change the battery 

when the battery energy is low. 

 

The device does not have an alarm function. Do not use the device in situations where alarms are required. 

 

The batteries  must be removed if the device is going to be stored for more than one month or else they  may 

leak. 

 

A flexible circuit connects the two parts of the device. DO NOT twist or pull on the connection. 

2 Overview 

The peripheral capillary oxygen saturation (SpO

2

) is the percentage of oxyhemoglobin (HbO

2

) in the total hemoglobin 

(Hb) in the blood, or the oxygen concentration in the blood. It is an important bio-parameter for  

respiration. For the purpose of measuring the SpO

2

  more easily and accurately, our company developed the Pulse 

Oximeter. The device can also measure the pulse rate simultaneously. 

The Pulse Oximeter features a small size, low power consumption, convenient operation and portability. The patient 

only needs  to put one of their  fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will 

show the measured value of Hemoglobin Saturation.  

2.1 Classification:  

Class II b, (MDD93/42/EEC  IX  Rule 10)

 

2.2 Features 

Operation of the device is simple and convenient. 

The  device  is small in size and  light in weight (total weight is about 50g including batteries)  making it

convenient to carry. 

The device has low power consumption. The two originally equipped AAA batteries can be operated

continuously for 20 hours. 

The device will automatically power off if no signal is received for 5 seconds. 

2.3 Major Applications and Scope of Application 

The Pulse Oximeter  can be used to measure human Hemoglobin Saturation and pulse rate and indicate the pulse 

intensity on the bar-display. The device is suitable for use in the home, hospital (ordinary sickroom), Oxygen Bar, and 

social medical organizations. 

 

The device is not suitable for use in continuous supervision for patients.

 

The problem of overrating will emerge if the patient is suffering from toxicosis caused by carbon monoxide. 

The device is not recommended to be used under these circumstances.

 

2.4 Environment Requirements 

Storage Environment 

 a) Temperature :-40

~+60

 

 b) Relative humidity 

:≤95%

 c) Atmospheric pressure :500hPa~1060hPa 

Operating Environment 

 a) Temperature :10

~40

 

 b) Relative Humidity 

:≤75%

 c) Atmospheric pressure :700hPa~1060hPa 

3 Principle and Caution 

3.1 Principle of Measurement 

Principle of the Oximeter: Principle of the Oximeter: for the calculation of blood oxygen level the device uses a 
formula to process the data according to Lambert Beer  Law and Spectrum Absorption Characteristics of Reductive 
Hemoglobin (Hb) and Oxyhemoglobin (HbO2) in glow & near-infrared zones.

Operation principle of the  Oximeter: Photoelectric Oxyhemoglobin Inspection Technology is applied together  with 

Capacity Pulse Scanning & Recording Technology so that two beams of light of different wavelengths can be focused 

onto the human nail tip through a perspective clamp finger sensor. A measured signal is obtained by a photosensitive 

element  and acquired  information will be shown on-screen through treatment in electronic circuits and 

microprocessors. 

Figure 1 Operating principle 

3.2 Caution 

1. The finger should be placed properly or else it may cause inaccurate measurement (see the illustration in Figure

5.) 

2. The SpO

2

  sensor should not be used at a location  or limb tied with an  arterial canal or blood pressure cuff,  or

receiving intravenous injection. 

3. Make sure the optical path is free from any optical obstacles like rubberized fabric. 

4. Excessive ambient light may affect the measuring result. This  includes fluorescent lamps, dual ruby lights, 

infrared heaters, direct sunlight etc.

5. Strenuous action of the patient or extreme electrosurgical interference may also affect accuracy.

6. The patient cannot use nail enamel or other makeup.

3.3 Clinical Restrictions 

1. As the measurement  is taken on the basis of arteriole pulse, substantial pulsating blood flow of the patient  is

required. For a patient with a weak pulse due to shock, low ambient/body temperature, major bleeding, or use of 

vascular contracting drugs, the SpO

2

  waveform  (PLETH) will decrease. In this case, the measurement will be 

more sensitive to interference. 

2. For patients with a substantial amount of staining dilution drugs (such as methylene blue, indigo green and acid

indigo blue), carboxyhemoglobin (COHb), methionine (Me+Hb) or thiosalicylic hemoglobin, and some with  an

icterus problem, the SpO

2

 determination by this monitor may be inaccurate.

3. Drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious

errors of SpO

2

 measurements.

4. As the SpO

2

 value serves as a reference value for judgment of anemic anoxia and toxic anoxia, some patients with

serious anemia may also report good SpO

2

 measurement. 

4 Technical Specifications 

1)

Display Format:

 OLED Display;

   SpO

2

 Measuring Range: 

0% - 100%; 

   Pulse Rate Measuring Range:

 30 bpm - 250 bpm; 

   Pulse Wave Display:

 columniation display and the waveform display. 

2)

Power Requirements:

 2×1.5V AAA alkaline batteries (or rechargeable batteries), adaptable  range: 2.6V~3.6V. 

3)

Power Consumption:

 Lower than 30mA.

4)

Resolution:

 1% for SpO

2

 and 1 bpm for Pulse Rate.

5)

Measurement Accuracy: 

±2% in stage of 70%-100% SpO

2

, and meaningless when stage being smaller 

TORONTEK-G64 Pulse Oximeter