55
MAINTENANCE
5
Insert the new illumination lamp straight and tighten securely the two fixing screws of the
illumination lamp.
6
Attach the lamp house cover by matching the projection at the bottom part of the lamp
house cover with the groove of the body cover.
If the lamp house cover is left unfixed, an error is displayed on the control panel and
operations, including photography, cannot be done.
REPLACING THE XENON LAMP
• The service life of the xenon lamp is approx. 10000 flashes.
Replace the xenon lamp if it is burned out or becomes whitened.
1
Turn the
OFF and unplug the power cord, then wait for more than 5 min-
utes for the natural electrical discharge.
2
Push the bottom of the lamp house cover and pull it forward to remove it.
3
Remove the xenon lamp set screw by turning it counterclockwise with a coin.
CAUTION
To avoid electric shock, be sure to turn the power switch off and
unplug the power cord before replacing the lamp.
CAUTION
To avoid electric shock, do not replace the xenon lamp with a new one
immediately after it goes off.
NOTE
To avoid whitening due to fingerprints, do not touch the lamp with bare
fingers.
NOTE
Since the lamp is not resistible to shocks, handle it with particular
care. Do not drop.
POWER SWITCH
Xenon lamp set screw