41
3.2.9
Buzzer
Set buzzer On or Off
The procedure for operation as follow;
1
Press [CHANGE] switch
2
Press [ROTATION] switch to select [ON] or [OFF].
Each time pressing the switch, ON and OFF is changed.
3
After selection, press [ENTER] switch.
3.2.10
Warm-up
Conduct the warm-up to stabilize the sensitivity of photo detector
Press [SHIFT] switch to start warm-up.
The warm-up runs for 60 minutes.
The RD-80SA starts the warm-up automatically after turn on power. When usage condition is
changed, Please conduct warm-up manually. The warm-up can be cancel by pressing [RUN]
switch.
* BUZZER ON/OFF *
O N
* PMT AGING *
SHIFT:START
During Aging
60 minutes