FUNCTION SETTING USING SETTINGS SCREEN
SETTING THE MAP OF ANALYSIS SCREEN LAYOUT BUTTON 2
The map to be placed in the Layout Change button of the Analysis Result screen can be
selected. Before shipment, the default setting is "Ocular Maps."
1
Tap the Set Condition button of the set item display "Map of analysis screen layout but-
ton 2" of "DISPLAY."
2
Tap the Setting Change button and select the applicable map.
3
Setting is complete.
• For map types and details, see "DESCRIPTION OF MAP LAYOUT" on page 63.
• It is not possible to set custom maps and pupillometry map to custom layout but-
tons.
SETTING THE MAP OF ANALYSIS SCREEN LAYOUT BUTTON 3
The map to be placed in the Layout Change button of the Analysis Result screen can be
selected. Before shipment, the default setting is "Corneal Maps."
1
Tap the Set Condition button of the set item display "Map of analysis screen layout but-
ton 3" of "DISPLAY."
2
Tap the Setting Change button and select the applicable map.
3
Setting is complete.
• For map types and details, see "DESCRIPTION OF MAP LAYOUT" on page 63.
• It is not possible to set custom maps and pupillometry map to custom layout but-
tons.
SETTING THE MAP OF ANALYSIS SCREEN LAYOUT BUTTON 4
The map to be placed in the Layout Change button of the Analysis Result screen can be
selected. Before shipment, the default setting is "Zernike Maps."
1
Tap the Set Condition button of the set item display "Map of analysis screen layout but-
ton 4" of "DISPLAY."
2
Tap the Setting Change button and select the applicable map.
3
Setting is complete.
• For map types and details, see "DESCRIPTION OF MAP LAYOUT" on page 63.
• It is not possible to set custom maps and pupillometry map to custom layout but-
tons.
SETTING CORNEAL IRREGULAR ASTIG.1 (VALUE 1<VALUE 2)
The Corneal Irregular Astig.1 of the IOL Selection map is set.
• In the following cases, the character color is changed based on the set threshold values of
Corneal Irregular Astigmatism 1 and Corneal Irregular Astigmatism 2:
• When the measurement value is not more than the set threshold value of Corneal Irregu-
lar Astigmatism 1
• When the measurement value is more than the threshold value of Corneal Irregular Astig-
matism 1 and not more than the threshold value of Corneal Irregular Astigmatism 2