48
3.6.1 Checking/Setting the Correction Factor
The values of the registered correction factors are displayed. If there are no factors registered,
"NO DATA" is displayed. "SET" is displayed underneath the number of the correction factor
currently being used.
[CHANGE]
: The displayed correction factor takes effects. When it is set, "SET" will be
displayed underneath the number of the correction factor.
When the value of the correction factor is not set, the input screen for
inputting the correction factor will be displayed.
Correction Factor Input Screen
☞
「
3.6.2 Inputting the Correction Factor
」
[ROTATION]
: Display the correction factor of next number.
[SHIFT]
: Each time the button is pressed, the display changes as shown below.
Reference data
→
Sample data
→
Correction factor data
→
…
- Reference Data -
The reference data from when the displayed correction factor was calculated
is displayed.
- Sample Data -
The sample data from when the displayed correction factor was calculated is
displayed.
Calculating the Correction Factor
☞
「
3.6.2 Inputting the Correction Factor
」
[ENTER]
: Returns to the previous screen.
・
Numbers such as "REF01" and "SMP01" that are displayed at the top left of the reference
data screen and sample data screen correspond to the number of the correction factor. In
other words, when the correction factor for K01 is calculated, the reference data for K01 is
REF01 and the sample data for K01 is SMP01.
K01
KX=
1.000E
+
00
SET
KY=
1.000E
+
00
KZ=
1.000E
+
00
C=SET
R=NEXT
S=PAGE
Memo
SMP01
x=
0.4476
y=
0.4074
L=
1.000E+02
C=SET
R=NEXT
S=PAGE
REF01
x=
0.4476
y=
0.4074
L=
1.000E+02
C=SET
R=NEXT
S=PAGE