31
SYSTEM DIAGRAM
Photography screen (Tomogram scan position: Manual adjustment)
Tap the fundus/anterior segment live image area in Area 2, and the system shifts to the scan position
adjustment screen. Tap the inside of the adjustment range, and you can change the scan position
without using the buttons. Changing the scan position is valid except "3D scan".
In "Line" scan
In "Radial" scan
[RESET] button
: Tap the [RESET] button. The change is discarded and the scan position is
reset to the initial status.
Scan position move
button
: Tap the upper, lower, right and left buttons on the fundus image to adjust the
scan position finely. Moving the scan position can be done in all scan pat-
terns except "3D".
Scan line
: Displays the scan position and direction.
Scan direction rotation
button
: Tap the upper button, and the scan direction rotates clockwise. Tap the
lower button, and the scan direction rotates counterclockwise. Rotation of
scan direction can be done in all scan patterns except "3D" and "Radial".
1234ABCD
Scan position move button
Scan line
Scan direction
rotation button
Scan position
adjustment range (blue)
[RESET] button
Scan possible range (red)