SOL-96.110.061 / 031221
TOPAZ Manganese
Configuration of the Analyzer
44
7.
Configuration of the Analyzer
Options Menu
To access the options menu, open the <TOOLS> tab from the main
screen and press the <Options>
button. Enter password “0712” if
necessary.
OPTIONS tab
Sleep mode
(min)
Set the time after which the display automatically turns off (tapping
turns it back on). If 0 seconds are set, the display will never turn off.
temp. setpoint
Setpoint of the regulated temperature in photometer.
Adjust this value only if instructed to do so by Seres OL customer
service.
pressure
setting
Setpoint of the water fault sensor.
Adjust this value only if instructed to do so by Seres OL customer
service.
Gain setpoint
Setpoint of the electronic gain.
Adjust this value only if instructed to do so by Seres OL customer
service.
JBUS slave ID
Set the JBUS slave ID to a number higher than 16.
Control mode
Select the appropriate control mode.
Status of relays, analog signal outputs and digital inputs in each
mode:
Setting
Relays
4/20 mA
Digital inputs Jbus
0 - Local
Activated
Activated
Frozen
(stop/start on
keyboard)
Reading
1 - Maintenance
Activated
Frozen
Frozen
Reading
2 - Jbus
Activated
Activated
Frozen
Reading,
Writing
3 - Remote control Activated
Activated
Activated
Reading