SOL-96.110.061 / 031221
TOPAZ Manganese
Product Description
9
2.1. Instrument Specifications
Power supply
Voltage:
Power consumption:
110–240 VAC
50/60 Hz
150 VA typical, 300 VA maximum
Environmental
conditions
Temperature:
Relative humidity:
Pollution level:
Installation category:
Maximum altitude:
5–40 °C
10% to 80%
2
II
2000 m
Do not use this instrument in presence of flammable liquids or gases.
Protect it from bad weather, humidity, corrosive substances and dust.
Process
connections
Sample inlet:
Sample outlet:
Sample outlet waste:
1/4” BSP F
soft tubing D INT 9 (single channel)
soft tubing D INT 19 (multi channel)
soft tubing D INT 12
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
Particle size:
min. 30 l/h (opt. 40 l/h)
5–40 °C
0.1–2 bar
pressure-free
<20 µm
Air cleaning
Inlet:
Required air flow:
6 mm tubing
50 l/h minimum of clean and dry air
Air cleaning can be used to protect the analyzer from a corrosive
environment.