Troubleshooting
47
8.
Troubleshooting
This chapter provides some hints to make troubleshooting easier.
For detailed information on how to replace or clean parts refer to
If you need help please contact your local distributor. Please have
the serial number and the diagnostic values (see
) ready.
8.1. What To Do If...
Instrument is not starting
Application is not launching
Instrument shows error message “Water default”
although there is sample flow
Possible cause
Corrective action
No power on the socket
– Check mains power supply.
Power supply defective
– Check output voltage of the
internal 24-volt power supply.
Panel PC not powered
– Check connections at the back
of the panel PC.
Possible cause
Corrective action
The instrument starts, but
the panel PC does not.
– Check connections at the back
of the panel PC.
Possible cause
Corrective action
Filing solenoid valve (SVB)
does not switch
– Check that the solenoid valve is
plugged in correctly.
One or more sample tubes
are blocked or pinched.
– Check condition of all sample
tubes (inside and outside).