69
Resistance
Rate
Range
(1)
Resolution
Full Scale
Reading
Test Current
Accuracy
(1 year)
Slow
5
00.00Ω
5
.0000 kΩ
5
0.000kΩ
5
00.00 kΩ
5
.0000MΩ
5
0.000MΩ
10mΩ
100mΩ
1Ω
10 Ω
100 Ω
1kΩ
510.00
5.1000
51.000
510.00
5.1000
51.000
0.5 mA
0.45 mA
45μA
4.5μA
450nA
45nA
0.10
%
+0.01
%
(2)
0.10
%
+0.008
%
(2)
0.10
%
+0.008
%
(2)
0.10
%
+0.008
%
0.15%+0.008
%
0.3
%
+0.01
%
(1)
In order to eliminate the noise interference, which might be induced to the test leads, it is
recommended to use a shielded test cable for measuring resistance above 100kΩ.
(2)
Using REL function
Maximum Input Protection: 1000VDC or 750VAC for all ranges.
Open circuit voltage: Maximum voltage is 5.5V dc
Continuity
Range
Resolution
Full Scale
Reading
Test Current
Accuracy (1 year)
23
℃
±5
℃
5
00Ω
100mΩ
999.9
0.5mA
0.1
%
+0.04
%
Maximum Input Protection: 1000VDC or 750VAC for all ranges.
Open circuit voltage: <5.5V DC
Test Current: around 0.5 mA DC
Threshold value:
2% of range
Diode
Rate
Range
Resolution
Full Scale
Reading
Test Current
Med
2.0000
V
100μV
2.3000V
0.5mA
Frequency
ACV
Range
Frequency
Range
Resolution
Full Scale
Reading
Accuracy
(2)
Input Sensitivity
(Sine Wave)
100mV
to
750V
5~10 Hz
10~100Hz
100~100 kHz
100k ~1MHz
(1)
100μHz
1 m Hz
10m Hz
10Hz
9.9999
99.999
999.99
999.99
0.05
%
+0.02
%
0.01
%
+0.02
%
0.01
%
+0.008
%
0.01
%
+0.008
%
200mV rms
300mV rms
300mV rms
500mV rms