■
3-10
3.2.2 Measuring refractive power and corneal curvature radius (Ref-
Keratometer measurement)
When using the data taken by this instrument to select
intraocular lenses, thoroughly determine the selection by also
examining cataract surgery methods and other inspections. If
incorrect measurement data is used to select intraocular lenses,
further surgery might be required.
When using the data taken by this instrument for refractive
correction surgery, thoroughly determine the selection by also
examining surgery methods and other inspections. Refractive
correction
surgery
conducted
according
to
incorrect
measurements or analysis results may result in further surgery
or severe complication such as keratoectasia.
Note that if diopter is used as the measurement unit for the radius
of corneal curvature, the measured value may differ depending
on the cornea equivalent refractive index. If an incorrect value is
used to select intraocular lenses, further surgery might be
required.
The corneal irregular astigmatism display function is a simple
measurement using keratometer measurement. This function
does not detect all cases of corneal irregular astigmatism.
Measurement itself or correct measurement may not be
performed for some corneal shapes, such as if there is a partial
problem on the corneal shape. Because the corneal irregular
astigmatism display function requires more information than
keratometer measurement, corneal irregular astigmatism may not
be measured even though keratometer measurement can be
performed.
Summary of Contents for MR-6000
Page 2: ......
Page 24: ... 1 18 This page is intentionally left blank ...
Page 30: ... 2 6 This page is intentionally left blank ...
Page 46: ... 4 2 This page is intentionally left blank ...
Page 56: ... 6 4 This page is intentionally left blank ...
Page 58: ... 7 2 This page is intentionally left blank ...
Page 67: ...8 9 This page is intentionally left blank ...
Page 68: ......
Page 70: ...0120 ...