ZT-1 User’s Manual – Operation
DOC #18172 Page 26
4) Install Shunt Resistor provided with each load cell, enclosed in calibration data report
plastic cover.
5) Balance the load cell to zero.
•
Adjust the ZERO pot until digital voltmeter shows 0.000 volts.
Note: Measure output voltage with a digital voltmeter across the OUT and COM
terminals (see Figure 13).
6)
•
Set the mode switches to AZ OFF, PRB, PEAK, and ZERO.