9. MAINTENANCE 

 

a) 

UV radiation is harmful  to eyes and skin.  It is  forbidden to directly observe UV 
lamp (radiator) during operation. 

b)  It is recommended to check cleanness of the quartz sleeve whenever the UV lamp is 

replaced. 

c)  If  any  sediment  is  noticed  on  the  surface  of the  quartz  sleeve,  it  is  recommended  to 

immediately clean the sleeve.  Negligence of this recommendation can cause drop in 
efficiency of sterilization.  

d)  Any time the sleeve is cleaned or replaced it is required to 

 

unconditionally replace the O-ring creating the seal around the sleeve.

 

e)  Inlet  and  outlet  valves  shall  be  opened  slowly  to  prevent  the  rush  of  fluid  hitting  and 

damaging the quartz sleeve. 

f)  There is no need for minimal flow through the device as it is designed in such a way 

that there may be no flow through the chamber with no damage to the sterilizer. 

When 

there is no flow of water 

through the chamber can heat up to approx. 85°C that is 

when turning off radiators occurs. Sterilizer is resumed to work after reaching 
temperature of approx. 65°C.

 

g)  UV  sterilizer  should  work  constantly  -  his  long-term  exclusion  may  cause  secondary 

infection in the installation, which is very difficult to remove. 

h)  It is recommended that before the UV sterilizer a filter is installed. 
i)  Water  not  fit  for  consumption  may  require  treatment  i.e.  deironing,  softening  or 

filtering. If any doubt, please contact your dealer. 

j)  In  case  of  high  turbidity  and  waste  water,  the  choice  of  equipment  and  efficiency 

should be made by the dealer.  

k)  Sterilizers must be sized for the maximum momentary flow of water (matching them to 

the daily flows may result in incomplete disinfection) 

l)  Avoid frequent turning on and off the equipment. One on-off cycle lowers the UV lamp 

life by about 10-20 hours.  

m) Sterilizers  should  not  be  installed  in  such  a  way  that  their  work  was  dependent  on 

other devices, e.g. pumps, pressure-water. 

n)  Flow  rate  through  the  device  can  be  increased;  however,  this  lowers  the  delivered 

dosage and lowers the efficiency of disinfection. 

o)  The lamp needs about 2 minutes to reach full capacity at temperature of water about 

12-

16 °C.  If the water is at temperature 5-12°C, the time might be about 3-5 minutes. 

p)  If the sterilizer is left idle for longer period of time or it can freeze, we recommend that 

the water be drained from the chamber.  The chamber is provided with drain valve. 

q) 

Bioassay  samples  shall  be  collected  into  sterile  opaque  containers  as  not  to 
expose  the  sample  to  the  light  to  avoid  photo  reactivation  of  microorganisms, 
which can rebuild their damaged DNA in presence of light

r)  In  case  the  red  indicator  light  is  on,  replacement  of  the  UV  lamp  is  necessary.    The 

part number can be found on transparent plastic lamp hood placed over the muff. 

s)  Even short lived surge, to above 254V, can damage the electronic ballast driving the 

UV lamps. 

Summary of Contents for AP-POOL-X 1

Page 1: ...Made in Poland WATER UV STERILIZERS AP POOL X SERIES MANUAL Patent No 204935 December 2019 WATER UV STERILIZERS...

Page 2: ...AP POOL X16 MODELS 10 5 INSTRUCTION FOR INSTALLATION AND OPERATION 11 5 1 CHAMBER ASSEMBLY 11 5 2 QUARTZ SLEEVES ASSEMBLY 13 5 3 UV LAMPS AND ELECTRICAL ENCLOSURE ASSEMBLY 14 5 4 MOUNTING TEMPERATURE...

Page 3: ...led and grounded in accordance with this manual and the local electrical code To guard against electrical shock unplug the device every time work is to be done on it GROUNDING Operation of the device...

Page 4: ...rated for the device New features allow continuing disinfection even without the flow of water through the chamber Sterilizers do not require control of flow and temperature of water by the user The s...

Page 5: ...ater Sterilizer uses special low pressure UV lamp to produce UV radiation of wavelength of 254 nm and 185nm which causes photochemical reaction damaging DNA of microorganisms leading to death or steri...

Page 6: ...CHAMBER Chamber Material Stainless Steel AISI 316 316L Finish Polished Powder coating Safety Class IP 66 Operating Pressure 3 5 bar 0 35MPa Operating Temperature Recommended 0 5 50 C Quartz Sleeve Gu...

Page 7: ...50 kg STERILIZER CHAMBER Chamber Material Stainless Steel AISI 316 316L Finish Polished Powder coating Safety Class IP 66 Operating Pressure 3 5bar 0 35MPa Operating Temperature Recommended 0 5 50 C Q...

Page 8: ...is used to empty the chamber from fluid The power supply system is mounted in the electrical enclosure and connected with the sterilizer by hi potential wires The enclosure houses besides the power s...

Page 9: ...ter removing the bridge it is possible to connect remote controlled switch to the ON OFF terminals caution voltage 230V CONTACTS 1 2 CLOSED sterilizer on CONTACTS 1 2 OPEN sterilizer off BALAST ON OFF...

Page 10: ...amp damage Alarm on Power cut off 4 5 CONNECTION OF SOLENOID VALVE IN AP POOL X2 AP POOL X16 MODELS There is possible to connect solenoid valve max 2A to the terminals 1 2 neutral on the electronic pl...

Page 11: ...ges without appropriate support f Upstream of the intake install potable water filter minimum rating 0 1 mm recommended 0 05 mm Water shall not contain more than 0 3mg Fe and 0 1 Mn g Install cut off...

Page 12: ...NUAL AP POOL X SERIES 12 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 Vertical Forbidden Supports Horizontal Supports Supports Horizontal Horizontal For...

Page 13: ...all not break the sleeve c Fill the chamber under full pressure and check the seal tightness ATTENTION Tightening the muffs with other tools may cause quartz sleeves breakage O RING STERILIZER CHAMBER...

Page 14: ...ot done as required the UV lamps may break after turning on the device LAMP HOOD STRAIN RELIEF PLACE SOCKET ON LAMP END WITH FOUR PINS UV LAMP SOCKET CENTERING RING UV LAMP STRAIN RELIEF INSERT UV LAM...

Page 15: ...d shut off valve shall be connected through terminals 1 2 Solenoid shut off valve serves the purpose of preventing unsterilized fluid entering the installation downstream of the sterilizer in the case...

Page 16: ...isplay An intermittent sound signal UV lamp reached its end of life time HOLD 3 s PRZYTRZYMA 3 s Hold left button more than 3 seconds to silence the sound alarm max 2 times Blinking number 0 on the di...

Page 17: ...n on the power switch while still holding the right button for more than 10 seconds f Short sound signal will confirm resetting the work time counter 8 LEGEND FOR INDICATOR LIGHT ON ENCLOSURE Lighted...

Page 18: ...led i Water not fit for consumption may require treatment i e deironing softening or filtering If any doubt please contact your dealer j In case of high turbidity and waste water the choice of equipme...

Page 19: ...breaker C10 AP POOL X6 Circuit breaker C13 AP POOL X8 Circuit breaker C16 AP POOL X10 Circuit breaker C20 AP POOL X12 Circuit breaker C16 AP POOL X16 Circuit breaker C16 10 INSTRUCTION FOR REPLACEMEN...

Page 20: ...uartz sleeve g Slide the sleeve out of guides and nipple of the sterilizer h Every time the UV lamp is replaced when it is necessary to clean the quartz sleeve use common glass cleaners and then insid...

Page 21: ...s in item 1 1 Procedures in the event of failure in multi radiator devices Description The alarm indication LED RED on one of the UV radiators lights up in the control cabinet and there is constant au...

Page 22: ...is NOT working Alarm system signals failure red LED Damaged electrical ballasts Alarm system signals proper work green LED Damaged electrical ballasts and or damaged alarm system UV lamp is working A...

Page 23: ...mm DN D mm a mm b mm c mm APP X1 276 220 1700 1470 72 50 3 220 325 250 120 APP X2 320 220 1735 1456 110 DN100 220 400 400 200 APP X3 375 250 1735 1430 125 DN125 256 400 600 250 APP X4 420 340 1740 13...

Page 24: ...01 2 pcs 4 pcs 6 pcs 8 pcs 10 pcs 12 pcs 16 pcs 20 pcs 24 pcs 32 pcs 4 O RING MUFF 01 04 01 1 pcs 2 pcs 3 pcs 4 pcs 5 pcs 6 pcs 8 pcs 10 pcs 12 pcs 16 pcs 5 WASHER 01 05 1 pcs 2 pcs 3 pcs 4 pcs 5 pcs...

Page 25: ...TMA MANUAL AP POOL X SERIES 25 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 16 HYGIENIC CERTIFICATE...

Page 26: ...TMA MANUAL AP POOL X SERIES 26 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 17 DECLARATION OF CONFORMITY...

Page 27: ...TMA MANUAL AP POOL X SERIES 27 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 18 GRAFICAL MOUNTING INSTRUCTION...

Page 28: ...TMA MANUAL AP POOL X SERIES 28 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 Assemble of AP POOL X sterilizer...

Page 29: ...TMA MANUAL AP POOL X SERIES 29 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 Assemble of UV lamps in AP POOL X chamber 2 3 4 5 1...

Page 30: ...all meet all specifications of performance as stated in the Manual for the period of 36 months from the date of purchase covering the stainless steel chamber 24 months from the date of purchase coveri...

Page 31: ...e is used for purposes and in a manner inconsistent with the recommendations applicable in the manual data sheet warranty card Damage caused by unforeseen natural phenomena i e floods frost storms hur...

Page 32: ...POOL X SERIES 32 TMA Bia ostoczek 26 gm Zab ud w 15 592 Bia ystok Poland Tel 48 857431246 VAT EU PL5420008413 Warranty and post warranty repairs Receive date Type of failure and repair procedure End d...

Reviews: