Tissue-Tek

®

 Film

®

 

 

Automated Coverslipper

 

 

 

 

 

 

 

 
 

 
Ensure 

Quick Reference Guide

 

The  Tissue-Tek

®

  Film

®

  automated  coverslipper  is 

designed  to  coverslip  biological  specimens  that  are 
mounted  on  standard  25  x  75  mm  (1.0  x  3.0  inch) 
microscope  slides.  The use of a  special,  resin-coated, 
film and xylene eliminates the need for coverglass and 
mounting  media.

  The  instrument,  as  part  of  the 

histopathology  process,  is  intended  to  facilitate  the  in 
vitro  examination  of  human  and  animal  tissue  for 
morphology changes by a pathologist.

 

This Quick Reference Guide contains information about 
day-to-day  use,  maintenance  and  cleaning.  For  more 
detailed  information  we  refer  to  the  most  recent 
Operating Manual of this instrument. 

All  users  should  be  trained  for  the  use  of  the  Sakura 
instruments.  The  Sakura  User  Training  is a  training  to 
ensure  that  new  and  existing  users  of  a  Sakura 
instrument have the knowledge and skills to operate and 
maintain the instrument in a proper way. 

 

Daily operations 

To  operate  the  instrument 

switch  on  the  “

POWER

” 

button located near the control panel. 

By turning on the power there will be a pre-operational 
check. 

Check the volume of xylene bottle and empty the waste 
bottle  if  necessary.  When  air  bubbles  are  in  the  line 

press “

PRIME

” to remove them. 

Ensure the Film is placed in the instrument as indicated 
in  the  image  on  the  right.  Follow  the  numbered 
positioning  spools  and  rollers  until  it  has  reached 
number  eight. 

Now  Press  “

FILM

”  to  feed  the  film 

automatically through the system until you can grab it at 
the end. 

When  the  Film  is  properly  fed  press  the  cutter  blade 
down to signal the system it is ready for use. Remove 
any access Film cut off in this way. 