Timberline Instruments TL-105-D Instruction Manual Download Page 1

 

 

 

 

 

TL-105-D COLUMN HEATER 

 

I

NSTRUCTION 

M

ANUAL

 

 
 

 
 
 
 
 

 

 
 
 
 
 
 
 
 

Timberline Instruments Inc. 

 

1880 S. Flatiron Ct., Unit I 
Boulder, Colorado 80301 
 
 

Ph: (303) 440-8779 
Fx: (303) 440-8786 
 
 

[email protected] 
www.timberlineinstruments.com

Summary of Contents for TL-105-D

Page 1: ...5 D COLUMN HEATER INSTRUCTION MANUAL Timberline Instruments Inc 1880 S Flatiron Ct Unit I Boulder Colorado 80301 Ph 303 440 8779 Fx 303 440 8786 info timberlineinstruments com www timberlineinstrument...

Page 2: ...escriptions 3 Thermal Fuse Protection 3 Electrical Fuse Protection 3 MOBILE PHASE HEAT EXCHANGER 4 INJECTION OR SELECTION VALVE 4 INSTALLATION 5 CONNECTING THE TL 105 D 5 INSTALLING AN INJECTION OR SW...

Page 3: ...es as temperature increases Improved peak shape by reducing on column silanol effects Improved ion exchange efficiency Lower backpressures by to reduction in mobile phase viscosity Selectivity changes...

Page 4: ...radients in the column When the mobile phase temperature is lower then the column temperature the fluid temperature in the heated column will be higher near the wall then it is in the center This will...

Page 5: ...ll reside in the same thermal environment as the heat exchanger and columns it is advised that the heat exchanger be plumbed into the system before the valve Placing the heat exchanger before the inje...

Page 6: ...ment will drain from the back of the compartment to the bulkhead on the right side of the oven near the lower back corner Some corrosive or harsh organic liquids are incompatible with the drain materi...

Page 7: ...alloy Heating Elements Etched Foil Silicon 150W Temperature Range TL 105 D Application dependent Tubing Material 316 SS or Hastelloy Temperature Stability 0 1 C Operating Voltage 120 VAC or 240 VAC Mu...

Page 8: ...ll shipments are made F O B Boulder CO Instruments and major components will be packed and shipped via surface unless otherwise requested Supplies and or replacement parts are packed and shipped via U...

Page 9: ...rns without this number will be returned to you collect Be careful to address the shipment to the street address Shipper will not deliver to our post office box numbers We require that you prepay ship...

Page 10: ...4 C column compartment will drain 6 corrosive or harsh organic liquids 6 D drain 6 H heat exchanger 4 I injection valve 4 5 P peak broadening 4 pre column heat exchanger 4 S set temperature 6 stabilit...