9.4
Arbitrary Waveform Generator
Output channel
1 analog, BNC
DAC resolution
14 bit @ 240 MS/s
Output range
-12 to +12 V (open circuit)
Amplitude
Range
0.12 V, 1.2 V, 12 V (open circuit)
Resolution
12 bit
Accuracy
0.4% of range
DC offset
Range
-12 V to +12 V (open circuit)
Resolution
12 bit
Accuracy
0.4% of range
Noise level
0.12 V
900
µ
V
RMS
1.2 V
1.3 mV
RMS
12 V
1.5 mV
RMS
Coupling
DC
Impedance
50
Ω
Overload protection
Output turns off when overload is applied. Instrument
will tolerate a short circuit to ground indefinitely.
System
Constant Data Size
Memory
Standard model
256 KiSamples
XM option
64 MiSamples
Operating modes
Continuous, triggered, gated
Maximum sampling rate
Depending on model
model 540, Model 530
240 MS/s
model 530
240 MS/s
model 220
200 MS/s
model 110
100 MS/s
model 055
50 MS/s
Sampling source
Internal TCXO
Accuracy
0.0001 %
Stability
±
1 ppm over 0
◦
C to +55
◦
C
Time base aging
±
1 ppm per year
Waveforms
Standard
Sine, square, triangle, pulse, noise, DC
Built-in arbitrary
Exponential Rise and Fall, Sin(x)/x, Cardiac, Haver-
sine, Lorentz, D-Lorentz
36
Chapter 9
Summary of Contents for Handyscope HS5 series
Page 1: ...Handyscope HS5 User manual TiePie engineering...
Page 6: ...2 Chapter 1...
Page 17: ...Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated Introduction 13...
Page 18: ...14 Chapter 3...
Page 24: ...20 Chapter 4...
Page 28: ...24 Chapter 5...
Page 30: ...26 Chapter 6...
Page 32: ...28 Chapter 7...
Page 46: ...42 Chapter 9...