Specifications
9
To achieve rated accuracy, allow the instrument to settle for 20 min-
utes. When subjected to extreme temperatures, allow additional
time for internal temperatures to stabilize. Because of temperature
compensated calibration, the Handyscope HS5 will settle within
specified accuracy regardless of the surrounding temperature.
9.1
Acquisition system
Number of input channels
2 analog
CH1, CH2
BNC
Type
Single ended
Resolution
8, 12, 14, 16 bit user selectable
Accuracy
0.25% of full scale
±
1 LSB
Range
±
200 mV to
±
80 V full scale
Coupling
AC/DC
Impedance
1 M
Ω
/ 25 pF
Maximum voltage
200 V (DC + AC peak
<
10 kHz)
Maximum voltage 1:10 probe
600 V (DC + AC peak
<
10 kHz)
Bandwidth (-3dB)
at 75% of full scale input
Ch1
250 MHz
Ch2
100 MHz
AC coupling cut off freq. (-3dB)
±
1.5 Hz
Sure
Connect
Optionally available (option S)
Maximum voltage on connection
200 V (DC + AC peak
<
10 kHz)
Resistance measurement
Optionally available (option S)
Ranges
100 Ohm to 2 MOhm full scale
Accuracy
1%
Response time (to 95%)
<
10
µ
s
Sampling source
Internal
TCXO
Accuracy
±
0.0001%
Stability
±
1 ppm over 0
◦
C to +55
◦
C
Time base aging
±
1 ppm per year time base aging
External
LVDS, on auxilary connectors
Input range
10 MHz
Memory
Standard model
128 KiSamples per channel
XM option
32 MSamples per channel when measuring 2 channels
64 MSamples when measuring 1 channel
Specifications
33
Summary of Contents for Handyscope HS5 series
Page 1: ...Handyscope HS5 User manual TiePie engineering...
Page 6: ...2 Chapter 1...
Page 17: ...Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated Introduction 13...
Page 18: ...14 Chapter 3...
Page 24: ...20 Chapter 4...
Page 28: ...24 Chapter 5...
Page 30: ...26 Chapter 6...
Page 32: ...28 Chapter 7...
Page 46: ...42 Chapter 9...