117
Gentier Real-Time PCR System
Caution:
please do not use organic or strong detergent to clean the
instrument shell, which
may ruin the surface coating.
1.2 Clean Touch Screen
1
st
step:
switch off the instrument and unplug the power cord;
2
nd
step:
gently wipe the touch screen with a piece of dry, soft cloth to remove dust, oil or
fingerprints.
3
rd
step:
if the touch screen is still not clean, use a piece of damp, soft cloth that moistened with
low concentration isopropanol or ethanol to clean the touch screen in a gentle motion.
Prohibit:
do not spray the detergent directly on the touch screen, as malfunctions of the
electronics or may occur.
Caution:
do not use abrasive detergent or rough material, as they may scratch the touch
screen.
1.3 Clean Sample Block
1
st
step:
switch off the instrument and unplug the power cord;
2
nd
step:
open the top lid and clean the sample block surface with a piece of damp, soft cloth, and
if needed, please rinse the cloth with mild commercial detergent.
3
rd
step:
clean the sample block wells with the
degreased cotton swab moistened with ethanol
to ensure the firm contact and good thermal conductivity between the consumables and
the wall of block well.
Prohibit:
do not spray the detergent directly on the thermal cycler block, as malfunctions of
the electronics or may occur.
Caution:
do not close the top lid before the sample block is completely dry.