Tianlan TM-103N User Manual Download Page 1

 

 

TM-103N Optical Power Meter 

User Manual 

                                  V1.0 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for TM-103N

Page 1: ...TM 103N Optical Power Meter User Manual V1 0 ...

Page 2: ...ecifications 3 4 Layout 4 5 Function Keys 5 6 Operation 7 6 1 Turn on off 7 6 2 Absolute power measurement 7 6 3 Relative power measurement 7 6 4 User self calibration function 8 6 5 Auto off function 9 6 6 Replace battery 9 7 Maintenance 9 8 Trouble Shootings 10 ...

Page 3: ...measurement It can display fast tracking and real time measurement of optical power when output power changes Wide measurement range it can provide 80dB of optical power measurement range Six calibrated wavelength 850nm 1300nm 1310nm 1490nm 1550nm and 1625nm other wavelength can be provided according to customer s request Provide absolute optical power measurement and relative power measurement Re...

Page 4: ...2 Auto off function Backlight LCD display AAA batteries which is more convenient to carry User self calibration function ...

Page 5: ...GaAs Accuracy 3 10dBm 22 Resolution Linearity 0 1 Non linearity 0 01dBm Working temperature 10 50 Storage temperature 20 70 Relative humidity 90 30 connector Movable FC PC SC PC ST as Optional Power supply 3pcs AAA batteries DC power supply Optional accessory which require special order Support wide range power input work AC 90 240 Working hours 120 hours without backlight Weight g 160g Dimension ...

Page 6: ...4 4 Layout 1 Wavelength Display the wavelength such as 850nm 1300nm 1310nm 1490nm 1550nm and 1625nm 2 Power value Displays the current output power under absolute ...

Page 7: ...ower Press REF for 2 seconds to start relative test 9 DBm w Absolute power 5 Function Keys 1 ON OFF press ON OFF to turn it on And press it for 3 seconds to turn it off Under power on mode press this key shortly to activate or deactivate the 10 minute auto off function The default setting is auto off function ON when start the meter Operators can press ON OFF shortly to close the auto off function...

Page 8: ... absolute power value as REF value and show the relative power value in dB unit C Enter the relative measurement mode and save current reference value in the absolute measurement mode press REF for 2 seconds to enter the REF measurement mode and save the current value as REF value which will appears in Ref position Now the value in dB unit on the LCD and the vale means the difference value between...

Page 9: ...velength is not exactly same to the one in the power meter then choose a close one from the power meter For example 1300nm wavelength is near to 1310nm we can choose 1310nm to substitue 1300nm Press dBm W to choose the display unit The default wavelength is 1310nm and working mode is dBm After the above steps the LCD will show the test power value 6 3 Relative power measurement Set the wavelength ...

Page 10: ... review the current reference value 6 4 User self calibration function During the operation when operators need to calibrate the meter to be accordance with other meters So our product offers operators self calibration function the following steps will tell you how to do Take out the batteries and pull up the switch 1 to ON position Put the battery back to the battery compartment and turn on the u...

Page 11: ... the opposition of the ON and turn off the meter End the user self calibration operation 6 5 Auto off function The auto off is on when turning on the meter Press ON OFF to activate or deactivate this function 6 6 Replace battery If the battery level is very low please turn off the unit immediately and replace the batteries Take out the batteries if not in use for a long period of time like one mon...

Page 12: ...n Solution Faint LCD screen Lower power Replace battery No backlight Lower power Replace battery No display when turning on Lower power other Turn on again or replace battery Fail to turn on Lower power Replace battery No changing on LCD screen Lower power Replace battery In sensitive LCD screen Dirty or polluted connector Use the correct connector and clean it Warranty Period 12 months from the p...

Page 13: ...vide the warranty card 3 Alter and omit the warranty card 4 Wrong operation including the problem caused by human actors abnormal working environment and so on 5 Problems or damages are not caused by the products quality 6 Do not operate it according to this manual 3 For the following situation we do not repair it 1 Damaged seal label 2 Take apart it without permission 3 Products from other manufa...

Page 14: ...12 Packing List 1 TM 103N Optical Power Meter 1pc 2 Soft case 1pc 3 User Manual 1pc 4 Battery AAA transportation permit 3pcs 5 Quality Certificate 1pc ...

Page 15: ...al number Date of production Inspection Warranty Card Date of purchase Customer name Customer address Customer phone Distributor name Distributor phone Posted back to Fault unit return back to manufacture for repair purpose should come with the warranty card with detailed information ...

Page 16: ......