21
vsp62-de-100921
PIN 1
PIN 2
PIN 3
PIN 4
PIN 5
PIN 6
+ 15...30VDC
GND
out
V
3k
10k
3.3.2
Operation With Other Supply And Evaluation Units
The transducer can be operated with other customer related display units or
voltage supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Pin Description:
1
2
3
6
5
4
Socket
Pin 1:
Pin 2:
Pin 3:
Pin 4:
Pin 5:
Pin 6:
Type Hirschmann 6pole, male
Line-Terminating Resistor (3.0 kOhm to GND)
Signal Output: 0-10 V (Measurement Range: 1.5-8.5V)
Signal Ground: AGND
Power Supply VCC: 15-30 VDC
Power Supply Common: GND
Shield
Shield (Pin 6) and supply common (Pin5) must always be
grounded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage
the transducer.