37
vsh82-de-110318
Survey of Commands:
Command
Code
Data Type
Function
Type
T
STRING
read
(VSH82: VSH208)
Measurement Value
M
FLOAT
read
Degas
D, d
BOOLEAN
read D, write d
Setpoint
S, s
FLOAT
read S, write s
(VSH82: SP1 and SP2)
Gas-Correction-Factor
C, c
UNSIGNED INT
read C, write c
(VSH82: C1 for Pirani,
C2 for Bayard Alpert)
Mode Hot Kathode
I, i
BOOLEAN
read I, write i
Mode
Sensor Transition
W, w
UNSIGNED INT
read W, write w
Pressure Adjustment
j
FLOAT
write
Examples (for address-switch in position "1"):
Action
Telegram to
transducer
Answer telegram
from transducer
Read type
"001Te
C
R
"
"001TVSH208p
C
R
"
Read pressure measurement
"001M^
C
R
"
"001M260014K
C
R
"
(
2.6x10-6 mbar)
Activate degas
"001d1f
C
R
"
"001d1f
C
R
"
Deactivate degas
"001d0e
C
R
"
"001d0e
C
R
"
Read setpoint 2
"001S2V
C
R
"
"001S400016O
C
R
"
(
4.0x10-4 mbar)
Unlock setpoint 2
Set setpoint 2 to
4.2x10-4 mbar
"001s2v
C
R
"
"001s420016q
C
R
"
"001s2v
C
R
"
"001s420016q
C
R
"
Unlock gas-correction-factor 1
Set gas-correction-factor 1
to 1.20
"001c1e
C
R
"
"001c000120W
C
R
"
"001c1e
C
R
"
"001c000120W
C
R
"
Unlock adjustment for
atmosphere pressure
Adjust on atmosphere pressure
"001j1l
C
R
"
"001j100023a
C
R
"
"001j1l
C
R
"
"001j100023a
C
R
"
Before writing the parameters "c" (correction factor), "s" (setpoint) and "j"
(adjustment) you must activate them by sending the corresponding unlock
command!
Summary of Contents for VSH82
Page 1: ...VSH82 Vakuum Messumformer Vacuum Transducer Betriebsanleitung Operating Instructions...
Page 23: ...23 vsh82 de 110318...
Page 26: ...26 vsh82 de 110318 Konformit tserkl rung...
Page 48: ...48 vsh82 de 110318...
Page 51: ...51 vsh82 de 110318 Declaration of Conformity...
Page 52: ...vsh82 de 110318...