background image

 

 

 

 

 

 

Dokument 
IMA 203 

Rev 

Page 
1 (4)

 

 

Installation manual 

CF3-W-US, CF3-W-US-Disp

 

 

CO

2

 transmitter 

 

   

 

CF3-W-US    CF3-W-US-Disp 

 

General 

The IAQ-sensor product CF3-W-US (sensor for wall mounting) is designed to 
measure carbon dioxide (CO

2

) in rooms. Option - Disp displays the measured CO

2

 

value in ppm (parts-per-million) on the LCD. 

The units are designed for connecting to Direct Digital Control (DDC). The two 
parallel signal outputs OUT1 (0-10 V) and OUT2 (2-10 V or 4-20 mA) give linear 
signals corresponding to the measuring range.  

The OUT2 output also indicates the status by setting the output voltage to 1 V or 
current to 2 mA when the sensor self-diagnostics detects any error. 

To open the wall mounted housing 

 

Figure 1

 

Closed housing seen from the top and the side. The housing is opened by inserting a screw 

driver and pushing to the front side of the housing. The locking hooks will then be released. 

Front

Wall

Insert screw driver 
Push upwards to open

Front

Wall

Reviews: