3.3.3 Using The TCH002 Controller Hub
The TCH002 USB Controller Hub provides power distribution for up to six T-Cubes,
and requires only a single power connection (from a separate supply unit TPS006
supplied by Thorlabs). Further details are contained in handbook
ha0146T T-Cube
Controller Hub
, supplied with the unit.
3.3.4 Supply Voltage and Current Requirements
The TLS001 is fitted with a 2.1mm DC connector. Thorlabs offers a compact, 5V
single way power supply (TPS101) for use with a single laser source cube.
Alternatively, an external regulated 5V power supply can be used.
Warning
DO NOT PLUG A POWERED UP T-CUBE INTO THE TCH002 USB
CONTROLLER HUB. Always ensure that all power is disconnected from
the laser source T-Cube AND the hub before the T-Cube is plugged into
the hub. Failure to observe this precaution will seriously damage the T-
Cube unit and could result in personal injury.
Warning
Exceeding the power supply maximum voltage, or connecting a reverse
polarity voltage can cause permanent damage to the unit and could result
in personnal injury.
Table 3.1 Power Supply Ratings
Supply
Minimum
Maximum
Max Operating Current
+5V
+4.9V
+5.1V
500 mA