6 Communication with a PC
31
·
If the last command was a Steinhart-Hart parameter, then this method is selected.
6.3.4 Switching the I Share On / Off (INTEG)
Command
Explanation
Response Example
Programming
":INTEG ON"
Switches the I share on
":INTEG OFF"
Switches the I share off
Reading
":INTEG?"
Reads the status of the I share
[:INTEG ON<LF>]
[:INTEG OFF<LF>]
6.3.5 Reading the TEC Current (ITE)
Command
Explanation
Response Example
Programming
":ITE:MEAS <NR1>"
Programs I
TEC
to be the measurement value for “ELCH
1
)” on
position <NR1> (1...8) in the output string.
Reading
":ITE:ACT?"
Reads the actual TEC (or heater) current
[:ITE:ACT <NR3><LF>]
":ITE:MIN_R?"
Reads the minimum TEC current for I
TE
- ADC = 0000
[:ITE:MIN_R <NR3><LF>]
":ITE:MAX_R?"
Reads the maximum TEC current for I
TE
- ADC = FFFF
[:ITE:MAX_R <NR3><LF>]
":ITE:MEAS?"
Reads the position of the TEC current as measurement value in
the “ELCH” output string (
1....8
,
0
if not selected)
[:ITE:MEAS <NR1><LF>]
1
)
EL
ectrical
CH
aracterization