© 2019 Thorlabs
19
PM102x
4.2 Pin Assignment Output Connector
PM102
: PM102 has a universal DA-15 connector. The Pins are assigned as follows:
Pin
Function
Description
1
+5...36VDC
Pin for Alternative Power Supply with 5 VDC to 36 VDC
2
I/O 1
General Digital Input / Output port; 3 V logic (Output), 5 V tolerant for In-
put
3
I/O 2
General Digital Input / Output port; 3 V logic (Output), 5 V tolerant for In-
put
4
POWER
Software Configurable Analog Output; 0 V to 2.5 V
5
X POS
Software Configurable Analog Output for X-Position; -2.5 V to 2.5 V
6
Y POS
Software Configurable Analog Output for Y-Position; -2.5 V to 2.5 V
7
TxD
RS232; Connect to PC terminal RxD (PC DE-9 Pin2)
8
RxD
RS232; Connect to PC terminal TxD (PC DE-9 Pin3)
9
GND
Power Supply Ground
10
GND
Ground
11
GND
Ground
12
GND
Ground
13
GND
Ground
14
N.C.
NTC Thermistor Input; Measurement Range 0.1 k
W
to 100 k
W
15
GND
RS232 Signal Ground; Connect to PC Terminal Ground (PC DE-9 Pin5)
4.3 Pin Assignment of the Sensor Connector
The sensor connector allows the access of all Thorlabs “C-type" thermal power sensors. The
power meter interface uses the sensor calibration data, stored in the sensor connector, to cal-
culate the corresponding actual power levels.
Additionally, the PM102x is capable to support custom made detectors. Please read the follow-
ing instruction prior to connecting a self made sensor.
Note:
The PM102 features an amplifier after each Thermal Sensor Input port TH1, TH2, TH3, and
TH4 (Pins 4, 5, 8, and 9). All 4 amplifiers are always on the same range.
Warning
Pin 2 is uniquely used for the EEPROM Digital I/O (memory in Thorlabs sensor heads) and
MUST NOT be used. Connecting this Pin may cause malfunction of the PM102x.
Pin-out of the DE-9 female
connector (female)