ORIC
®
Piezo Stage Controller PDXC
Chapter 4: Panel Operation
Page 18
CTN015391-D02
4.3.1. Channel Selection
To select the channel in SMC mode, press the
“ENTER” button in main SMC mode screen to enter the menu. Then
follow below flow chart select channel between Channel 1 (CH1) and Channel 2 (CH2).
Figure 4.17 Select Channel in SMC Output Mode
Panel operation only allows the user to select one output channel in SMC output mode. The user can only move
one stage at a time. Moving two stages simultaneously would require software or command line operation.
4.3.2. Setting Velocity Value
In SMC output mode, the moving speed of the stage is defined by how many increments the stage took within a
second. This corresponds to how many pulses of driving signal are sent to the stage from the controller. In SMC
mode, the value of the speed can be set between 800 to 20000 steps (pulses) per second.
To set the velocity in SMC mode, press the
“ENTER” button at the main SMC mode screen to enter the menu. Then
follow the flow chart below.
Select SMC mode after power on
SMC mode main screen
Press “ENTER” to enter the menu screen
and choose “SELECT CHANNL”
Turn knob to select output channel
an
d press “ENTER” to validate
Press “ENTER” again to confirm selection