ORIC
®
Piezo Stage Controller PDXC
Chapter 4: Panel Operation
Rev B, September 23, 2021
Page 15
Figure 4.9 Save Parameters in D-Sub Output Mode
4.2.7. Homing
Homing is a necessary process for closed-loop operation to be accurately executed. We recommend homing every
time the stage is switched to closed-loop mode from open loop operation, or after power cycles.
The homing process scans
the stage across its entire travel range and looks for a “HOME” position to set as the
absolute reference position for the stage. All the stage movements are executed relative to that HOME position.
To execute the homing process, follow the flow chart below:
Press “ENTER” to enter the menu screen
and turn knob to toggle down the screen
until it shows “SAVE”
D-Sub mode main screen
Press “ENTER” to save all operation
parameters in D-SUB mode
Press “ENTER” again to confirm