Rev A
Apr
2020
Page 7
Chapter 4 Installation
4.3.2 Mounting to the Work Surface
The stage is shipped secured to the base plate with 3/16" hex, 1/4"-20 (imperial version) or 5 mm hex, M6 (metric version)
cap screws. Skip steps 1 through 4 if starting with the base plate already
attached.
Referring
to Fig. 4.1, proceed as follows:
1)
Turn the motor knob to position the moving carriage central in its range of travel, and ensure that the mounting holes in the
base are clearly vis
i
ble through the holes in the top plate.
2)
Fit the dowels supplied to the base plate (LNR50P
4
).
3)
Position the stage on the base plate, ensuring that the dowels locate correctly in the holes in the lower surface of the stage.
4)
Fit two bolts (M6 x 12
mm
or 1/4
"
-20 x 1/2
"
) through the holes in the top plate and into the base plate.
Tighten to
secure
the stage in place.
Fig. 4.1 Fitting the
B
ase
P
late
5)
Fit three bolts (M6 or 1/4
"
-20, not supplied) through each end of the base plate to fix the stage to the work
surface.
6)
To remove the base plate, reverse the procedure above.
1
4
1
2
5
3
Summary of Contents for LNR502