LFE1220W-US, -EU, and -UK Laminar Flow Enclosures
Chapter 3: Installation
Page 36
ETN056370-D02
3.7. Fitting the HEPA Fan Filter Unit (FFU)
3.7.1. Required Parts:
Code
Description
Qty
L
Roof Panel Outer
2
N
HEPA FFU (Fan Filter Unit)
2
AG
Barrel Plug
22
AH
Push Rivet
28
AL
Cable Gland
2
3.7.2. Fitting the HEPA FFU
39. Wire up the HEPA FFU (N). Please refer to the
MAC 10 Original VE5 FAN FILTER MODULE INSTALLATION
AND SERVICE MANUAL
(included) for more detailed instructions and for wiring diagrams.
To protect the electrical cable against chafing, a cable gland (AL) has been provided for installation in the
electrical entrance.
Risk of Electrical Shock
Installation work and electrical wiring must be done by qualified person(s) in accordance with all
applicable codes and standards.
Cables for wiring the Filter Unit are not supplied. These must be provided by the person responsible for
installation and must conform to any applicable local standards.
The unit must be installed only by qualified personnel who understand the hazards associated with using
high voltages and the steps necessary to minimize the risk of electrical shock.
Caution
Touching of the HEPA filter could damage it.
Never place a hand or tool on the filter.
Never lay the filter facing flat down on a surface. Always have filter on its side or back to protect it from
damage.