LFE1220W-US, -EU, and -UK Laminar Flow Enclosures
Chapter 3: Installation
Rev A April 2021
Page 9
2. The construction rails are assembled by sliding the head of low profile screws into the T-slots of the construction
rail. To enable the assembly, the spacing between the end of the construction rail and the underside of the screw
head must be set to a specific distance. For convenience, all assembly screws have been pre-fitted, but the exact
spacing must be set before assembly. An assembly tool (AK) has been provided for this purpose.
Set the spacing of the low profile screws in the ends of the RH and LH Longitudinal Supports
(A
) and (B
).
Using the 4mm T-handle hex driver (AJ), loosen the screws from their pre-installed position
just enough to insert the assembly tool (AK) under the screw heads. Tighten the screws lightly against the
assembly tool, then back off 1/4 turn to release the assembly tool.
Figure 2 Using the Assembly Tool
3. Using the 4mm T-handle hex driver (AJ) supplied, loosen the 3 screws holding the flange to the Lateral Filter
Support (C
).
Loosen just enough
for the flange to slide along the slot (<1 turn). Slide the flange to the left.
Figure 3 Sliding the Flange