DET02AFC Operating Manual — Fiber Input Si Photodetector 

Description: 

The DET02AFC is a ready-to-use, high-speed Si photodetector for use with FC/PC connectorized fiber optic cables 
in visible wavelength optical systems. The unit comes with an FC/PC bulkhead connector, detector, and 12V bias 
battery enclosed in a compact aluminum housing. The FC/PC connector provides easy coupling to fiber-based light 
sources. The output uses an SMA jack to minimize size and maximize frequency response. The bandwidth is 
1.2GHz and will operate over the spectral range of 400-1100nm. A NIR version, the DET01CFC, is also available 
and operates in the 800-1700nm spectral range. 

The DET02AFC includes one #8-32 tapped mounting hole with a 0.25

 mounting depth, while the DET02AFC/M has 

one M4 tapped mounting hole with a 6.35mm mounting depth. A 12V A23 battery is included. 

 

Specifications*: 

Detector: Si 

PIN 

 

Output Voltage

(Hi-Z): 0-10V 

Active Diameter:  Ø0.25mm 

 

Output Voltage

(50

): 0-410mV 

Wavelength Range (

λ)

:  400-1100nm 

 

Output:  SMA (DC coupled) 

Peak Wavelength (

λ

p)

:  730nm (typ) 

 

Dark Current

(I

D

):  50pA (500pA max) 

Peak Response [typ, 

(

λ

p

)]:  0.46A/W (typ) 

 

Shunt Resistance (R

sh

): 20G

 

Input:  FC/PC fiber connector 

 

Weight: 0.18lbs 

Rise Time (t

R

): 50ps 

(typ) 

 

Storage Temp:  0-40

°

Fall Time (t

f

): 250ps 

(typ) 

 

Operating Temp:  0-40

°

 Ball Lens Size:  Ø0.059

  (Ø1.5 mm) 

 

Battery: A23, 

12V

DC

, 40mAh 

Diode Capacitance (C

J

): 0.94pF 

(typ) 

 

NEP (730nm):  9.5x10

-15

W/

Hz (max) 

Min Load Resistance:  50

Ω

 

 

Bias Voltage (V

R

): 12V 

Damage Threshold:  18mW 

 

   

*All measurements performed with a 50

Ω

 load unless stated otherwise. 

1.

 

Into high impedance. Max voltage is dependent on battery voltage. New batteries typically show 12-13V, while drained 
batteries show 9V or less. 

2.

 

Calculated range based on peak responsivity and damage threshold. 

3.

 

Measured with specified bias voltage. 

 

 