6 Operation Instructions
6.1 Setup BC207 Series
This chapter describes the setup for the BC207UV(/M) and BC207VIS(/M).
1.
Provide stable mounting of the Beam Profiler as described in the chapter
.
2.
Download the Beam Software version 8.0 or higher through the software tab on the BC207
Series
it on your computer.
3.
With the filter cap still on the beam profiler, adjust the BC207 Series Camera Beam Profiler
such that its optical aperture will be exposed to the optical beam.
4.
Align the beam to be measured so that it is perpendicular to the front face of the Beam
Profiler.
5.
Be sure to operate the instrument within its specified
. Please regard the
output power of your light source.
a.
Choose a neutral density (ND) filter in the
which is suited for the operat-
ing
wavelength
(UV / VIS) of your light source.
b.
Select a filter attenuation which is adapted to the applied laser power. Always start
with the
maximum attenuation
to prevent damage to the beam profiler.
c.
If required, choose additional attenuation in front of the beam profiler to stay within
the specified power range.
6.
Minimize ambient light entering the Beam Profiler aperture.
7.
Remove the filter cap.
8.
Connect the camera using the supplied USB 3.0 cable to the PC as described in
.
9.
the Beam Software.
10. Switch on the light source.
Attention
Do not stick anything into the Beam Profiler aperture
!
The protective glass might be
scratched.
Prevent dust or other contamination from entering the aperture!
Keep beam power below the allowed limit to avoid damage to the instrument!
6.2 Measurement with the Beam Profiler
To achieve correct and reliable measurement results, it is recommended that these basic
guidelines are followed.
1. Follow the instructions described in
.
2. In the Beam Software:
.
of the chosen ND filter.
c. Perform the
in case of weak light sources compared to the
ambient power level.
Software Performance Optimization
As soon as a BC207 Series device is selected within the "Device Selection" panel, the mea-
surement starts in the continuous mode. It may be advantageous to stop the continuous mea-
surement for a detailed analysis of a beam profile captured with the last image. Also, user inter-
actions with the GUI will work more fluently when the continuous flow of image data is stopped.
7
13
164
5
13
14
7
27
28
29
28
Summary of Contents for BC207UV
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 10 Jan 2022 ...
Page 103: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 96 2022 Thorlabs ...
Page 133: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 126 2022 Thorlabs ...
Page 181: ...www thorlabs com ...